![Hill-Rom The Vest P205 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/hill-rom/the-vest-p205/the-vest-p205_user-manual_2134210016.webp)
10
The Vest
®
Airway Clearance System, Model 205
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
Full SPU Vest
NOTE:
It is recommended that a single layer of cotton clothing be worn beneath
the disposable garment.
1. Separate the front flaps of the Full SPU Vest at the Velcro®
fasteners.
2. Turn the Full SPU Vest so the front flaps will meet in front of the
patient. Then put the patient’s arms through the arm openings in the
Full SPU Vest.
3. With the Full SPU Vest deflated, use the Velcro® fasteners to attach
the front flaps of the Full SPU Vest to each other.
4. Make sure there is
approximately 3" to 4"
(8 cm to 10 cm) between
the bottom edge of the
front of the Full SPU Vest
and the patient.
Adjust the Velcro® fasteners at
the shoulder straps so the
bottom edge of the Full SPU
Vest is level with the top of the
patient’s hip bone.
Connect the Air Hoses
1. Slide one end of each Air
Hose into each Air Hose
Connector Port on the
front of the Air Pulse
Generator. Use a slight
twist motion as you push
the Air Hose to help keep
it in position.