![Hill-Rom The Vest P205 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/hill-rom/the-vest-p205/the-vest-p205_user-manual_2134210027.webp)
The Vest
®
Airway Clearance System, Model 205
21
User Manual (140643 REV 7)
PROPRIETARY AND CONFIDENTIAL DRAFT 19-DEC-11
c.
To adjust the
treatment Time,
press the right
Up
or
Down
arrow
buttons until the
prescribed
treatment Time
shows on the
screen. The treatment Time may be set between 1 and 60
minutes.
7. Press the
ON
button to advance to the next screen. This screen
shows the summary of the ramp program you have just created.
8. Press the
ON
button.
The disposable
garment will inflate.
9. Press the
ON
button
again to begin the
therapy. The screen
will show the word
Ramping during the
ramp portion of the
program. The total
treatment Time shows
as it counts down to zero.
NOTE:
Therapy settings
cannot
be changed during Ramping stage.
10. If it is necessary to pause the treatment during a session, do as
follows:
a.
Press the
OFF
button or Remote Control once. The unit will
stop the pulsations, but the settings will continue to show on
the screen.
NOTE:
When the unit is paused, the disposable garment will deflate.