SERVICE MANUAL

Century

Bed

From Hill-Rom

For Parts Or Technical Assistance

 man241

USA (800) 445-3720 Canada (800) 267-2337
International: Contact your distributor.

Product No. P1400

errata

man241-e1

Содержание Century+ P1400

Страница 1: ...SERVICE MANUAL Century Bed From Hill Rom For Parts Or Technical Assistance man241 USA 800 445 3720 Canada 800 267 2337 International Contact your distributor Product No P1400 errata man241 e1...

Страница 2: ...Century Bed Service Manual man241 Page i Century Bed Service Manual Revisions man241 Revision Letter Pages Affected Date Original Issue April 1999...

Страница 3: ...s a registered trademark of Hill Rom Inc Hill Rom is a registered trademark of Hill Rom Inc Loctite is a registered trademark of Loctite Corporation Inc Oilite is a registered trademark of Beemer Prec...

Страница 4: ...of sale to the original purchaser Hill Rom makes no other warranty express or implied and in particular makes no warranty of merchantability or fitness for a particular purpose Additional copies of th...

Страница 5: ...Revisions Page iv Century Bed Service Manual man241 NOTES...

Страница 6: ...ist 1 6 Chapter 6 General Procedures 1 6 Chapter 7 Accessories 1 6 Typographical Conventions 1 7 Introduction 1 8 Overview 1 8 Operating Precautions 1 8 Bed Positions 1 8 Specifications 1 12 Physical...

Страница 7: ...ation 1 17 General Operation 1 18 Automatic Contour Optional 1 18 Bed End Panels 1 18 CPR Release Optional 1 18 Battery Backup 1 19 Motor Assemblies 1 19 Thermals 1 19 Manual Operation 1 20 Head Funct...

Страница 8: ...tion Does Not Work 2 27 The Knee Down Function Does Not Work 2 35 The Bed Up Function Does Not Work 2 43 The Bed Down Function Does Not Work 2 52 The Knee Section Does Not Raise Beds with Automatic Co...

Страница 9: ...4 10 Adjustment 4 11 Battery Backup Circuit Board 4 12 Removal 4 12 Replacement 4 14 Adjustment 4 14 Battery Backup Cells 4 15 Removal 4 15 Replacement 4 17 Adjustment 4 17 Head or Knee Motor 4 18 Rem...

Страница 10: ...enburg or Reverse Trendelenburg Assembly 4 36 Removal 4 36 Replacement 4 38 Adjustment 4 38 Brake Cam Assembly Beds with Central Brake and Steer 4 39 Removal 4 39 Replacement 4 40 Adjustment 4 41 Brak...

Страница 11: ...ment 4 59 Adjustment 4 60 Trendelenburg and Reverse Trendelenburg Indicator 4 61 Removal 4 61 Replacement 4 61 Adjustment 4 61 Chapter 5 Parts List Warranty 5 3 Service Parts Ordering 5 5 Exchange Pol...

Страница 12: ...ls Cable Routing 5 50 SideCom Communication System Head Siderail Assembly 5 52 SideCom Communication System Head Siderail Entertainment Controls 5 56 SideCom Communication System Head Siderail Bed Con...

Страница 13: ...7 4 Removal 7 4 Adjustment 7 5 Architectural Bumpers P819 7 6 Installation 7 6 Removal 7 6 Adjustment 7 6 IV Rod P2217 7 7 Installation 7 7 Removal 7 7 Adjustment 7 8 Trapeze Support P846 or P846D 7...

Страница 14: ...ment and Adjustment Procedures 1 5 Chapter 5 Parts List 1 6 Chapter 6 General Procedures 1 6 Chapter 7 Accessories 1 6 Typographical Conventions 1 7 Introduction 1 8 Overview 1 8 Operating Precautions...

Страница 15: ...1 15 Regulations Standards and Codes 1 16 UL Classification 1 16 Model Identification 1 17 General Operation 1 18 Automatic Contour Optional 1 18 Bed End Panels 1 18 CPR Release Optional 1 18 Battery...

Страница 16: ...Chapter 1 Introduction Century Bed Service Manual man241 Page 1 3 1 Fluids 1 27 Water Mattresses 1 27 Lockout Switches 1 27 CPR Release 1 27 Warning and Caution Labels 1 31...

Страница 17: ...Chapter 1 Introduction Page 1 4 Century Bed Service Manual man241 NOTES...

Страница 18: ...n This manual contains seven chapters Chapter 1 Introduction In addition to a brief description of this service manual chapter 1 also provides a product overview Chapter 2 Troubleshooting Procedures R...

Страница 19: ...ocedure and illustrated parts lists Chapter 6 General Procedures Cleaning preventive maintenance and other general procedures are described in this chapter Chapter 7 Accessories A list of additional p...

Страница 20: ...rification The symbol below highlights a WARNING or CAUTION Figure 1 1 Warning and Caution A WARNING identifies situations or actions that may affect patient or user safety Disregarding a warning coul...

Страница 21: ...rating the Century Bed be sure that you have read and understand in detail the contents of this manual It is important that you read and strictly adhere to the aspects of safety immediately following...

Страница 22: ...Introduction Chapter 1 Introduction Century Bed Service Manual man241 Page 1 9 1 Figure 1 6 Hilow Positions m241_009...

Страница 23: ...Introduction Chapter 1 Introduction Page 1 10 Century Bed Service Manual man241 Figure 1 7 Trendelenburg Reverse Trendelenburg Positions m241_010...

Страница 24: ...Introduction Chapter 1 Introduction Century Bed Service Manual man241 Page 1 11 1 Figure 1 8 Automatic Contour Position Foot end Head end m241_137...

Страница 25: ...in relation to the head section in order to provide the occupant with easy access regardless of the degree of inclination Feature Dimension Height in high position floor to sleep deck 30 1 4 76 8 cm H...

Страница 26: ...ction will operate as long as the switch is depressed and will stop once the switch is released The switches are labeled with pictures that indicate their function Siderail switches are fixed in relat...

Страница 27: ...prevent the patient from operating the head knee or hilow functions Lockouts are only available at the attendant control console and are used when necessary for patient safety Activation of the knee l...

Страница 28: ...length and two foot siderails as standard equipment The siderails can be retracted and stored under the sleeping surface frame Each head siderail contains head knee and hilow controls and can be fitte...

Страница 29: ...cifications Chapter 1 Introduction Page 1 16 Century Bed Service Manual man241 Regulations Standards and Codes UL Classification Class grounded equipment per UL 544 CSA risk class 2G per CA per CSA 22...

Страница 30: ...pter 1 Introduction Century Bed Service Manual man241 Page 1 17 1 Model Identification See table 1 2 on page 1 17 for Century Bed model identification Table 1 2 Model Identification Model Number Descr...

Страница 31: ...ure automatically raises the knee section of the bed up to 15 as you raise the head section The automatic contour feature helps prevent patients from sliding to the foot end of the bed when the head s...

Страница 32: ...ad section Battery Backup The battery backup allows all the motor functions of the bed to operate without external power This feature enables the caregiver or patient to adjust the bed during transpor...

Страница 33: ...ssembly B from the head end of the bed Figure 1 9 Head Function Manual Operation WARNING Unplug the bed from its power source and disconnect the motor before inserting the IV rod for manual operation...

Страница 34: ...d clockwise or counterclockwise to raise or lower the head function as desired WARNING Remove the IV rod before plugging the bed into its power source Failure to do so could result in personal injury...

Страница 35: ...can raise or lower the head section using the CPR release at the head end of the bed Depress the CPR release button A and pull out on the handle B see figure 1 10 on page 1 22 Manually raise or lower...

Страница 36: ...Manual Operation WARNING Unplug the bed from its power source and disconnect the motor before inserting the IV rod for manual operation Failure to do so could result in personal injury or equipment d...

Страница 37: ...unction Manual Operation You can raise or lower the knee section using a Hill Rom P2217 IV rod A as a crank see figure 1 12 on page 1 24 Access the knee drive B through the hole in the back of the kne...

Страница 38: ...end of the bed and into the hole in the knee drive motor casing 4 Turn the IV rod clockwise or counterclockwise to raise or lower the knee function as desired WARNING Remove the IV rod before plugging...

Страница 39: ...ice personnel to have their entire body below the sleep surface and within the confines of the bed Unplug the bed from its power source prior to cleaning or servicing it If service personnel need to g...

Страница 40: ...ch stress on the motor drives Generally the patient s weight plus the water mattress weight will exceed the recommended bed capacity More importantly water mattresses can rupture which would allow lar...

Страница 41: ...age WARNING Powered bed mechanisms can cause serious injury Operate the bed only with persons clear of mechanisms WARNING Only facility authorized maintenance personnel should troubleshoot the Century...

Страница 42: ...bed frame before working on the bed Failure to do so could result in personal injury or equipment damage SHOCK HAZARD Unplug the bed from its power source Failure to do so could result in personal inj...

Страница 43: ...revent component damage CAUTION Wear an antistatic strap when handling electronic components Failure to do so could result in component damage CAUTION Place the removed P C boards in antistatic protec...

Страница 44: ...Warning and Caution Labels Chapter 1 Introduction Century Bed Service Manual man241 Page 1 31 1 Warning and Caution Labels Figure 1 13 Warning and Caution Labels m241_015...

Страница 45: ...Warning and Caution Labels Chapter 1 Introduction Page 1 32 Century Bed Service Manual man241 NOTES...

Страница 46: ...es Not Work 2 11 The Head Down Function Does Not Work 2 19 The Knee Up Function Does Not Work 2 27 The Knee Down Function Does Not Work 2 35 The Bed Up Function Does Not Work 2 43 The Bed Down Functio...

Страница 47: ...Chapter 2 Troubleshooting Procedures Page 2 2 Century Bed Service Manual man241 NOTES...

Страница 48: ...to isolate or identify a problem and to verify the repair after completing each corrective action replacing or adjusting a part seating a connector etc Perform the Final Actions after the Function Che...

Страница 49: ...nts a common digital VOM The three basic electrical functions that you will test are alternating current AC direct current DC and ohms resistance Figure 2 1 Volt Ohm Meter VOM Figure 2 1 on page 2 4 d...

Страница 50: ...blem The problem can be duplicated Yes No Go to Function Checks on page 2 5 3 The problem is a result of improper operator action Yes No Go to Function Checks on page 2 5 4 Perform the Function Checks...

Страница 51: ...g properly perform the following function checks Otherwise go to Final Actions on page 2 7 6 Press the head up switch and activate the head section The head section raises when the head up switch is a...

Страница 52: ...d automatic contour function The head section and knee section lower when the head down switch is activated Yes No Go to RAP 2 9 14 When the bed is in the high position activate the Trendelenburg cont...

Страница 53: ...Final Actions Chapter 2 Troubleshooting Procedures Page 2 8 Century Bed Service Manual man241 NOTES...

Страница 54: ...ise go to step 4 4 Check circuit breaker on the transformer box at the head end of the bed The circuit breaker good Yes No Reset the circuit breaker If this solves the problem go to Final Actions on p...

Страница 55: ...elay on the control board see figure 3 6 on page 3 6 The relay closes when any of the four bed functions are activated Yes No Replace the control board see Control Board on page 4 5 If this solves the...

Страница 56: ...Go to step 19 2 Perform the following steps a Set your VOM to measure volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure to use caut...

Страница 57: ...e figure 2 3 on page 2 12 Figure 2 3 Head Motor Location b Set your VOM to measure volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure...

Страница 58: ...Failure to use caution could result in serious electrical shock injury b At P1 measure the voltage between common pin 1 and 32V pin 3 see figure 3 4 on page 3 5 c The voltage is 32V AC 3V Yes No Go to...

Страница 59: ...to step 27 15 Perform the following a Set your VOM to read volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure to use caution could re...

Страница 60: ...rail with the inoperative head up function e In the cable P17 or P18 check the resistance between VCC common pin 7 and the head up function pin 3 see figure 3 1 on page 3 3 and figure 3 4 on page 3 5...

Страница 61: ...isconnect the battery backup Failure to do so could result in personal injury or equipment damage b If applicable disconnect the battery backup see Battery Backup Box on page 4 8 c Open the siderail w...

Страница 62: ...ts the test port P200 see figure 2 2 on page 2 11 to the siderail with the inoperative head up function at the siderail end of the cable e Check the resistance in the switch see figure 3 3 on page 3 4...

Страница 63: ...tion at the siderail end of the cable e Check the resistance in the switch see figure 3 3 on page 3 4 and figure 3 4 on page 3 5 f While holding the VOM probes in place depress the head up switch g Th...

Страница 64: ...o Go to step 19 2 Perform the following steps a Set your VOM to measure volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure to use cau...

Страница 65: ...ee figure 2 5 on page 2 20 Figure 2 5 Head Motor Location b Set your VOM to measure volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failur...

Страница 66: ...Failure to use caution could result in serious electrical shock injury b At P1 measure the voltage between common pin 1 and 32V pin 3 see figure 3 4 on page 3 5 c The voltage is 32V AC 3V Yes No Go t...

Страница 67: ...to step 27 15 Perform the following a Set your VOM to read volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure to use caution could re...

Страница 68: ...il with the inoperative head down function e In the cable P17 or P18 check the resistance between VCC common pin 7 and the head down function pin 4 see figure 3 1 on page 3 3 and figure 3 4 on page 3...

Страница 69: ...OCK HAZARD Disconnect the battery backup Failure to do so could result in personal injury or equipment damage b If applicable disconnect the battery backup see Battery Backup Box on page 4 8 c Open th...

Страница 70: ...embly on page 4 52 d Disconnect the cable that connects the test port P200 see figure 2 4 on page 2 19 to the siderail with the inoperative head down function at the siderail end of the cable e Check...

Страница 71: ...erail with the inoperative head down function at the siderail end of the cable e Check the resistance in the switch see figure 3 3 on page 3 4 and figure 3 4 on page 3 5 f While holding the VOM probes...

Страница 72: ...Go to step 19 2 Perform the following steps a Set your VOM to measure volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure to use caut...

Страница 73: ...e figure 2 7 on page 2 28 Figure 2 7 Knee Motor Location b Set your VOM to measure volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure...

Страница 74: ...Failure to use caution could result in serious electrical shock injury b At P1 measure the voltage between common pin 1 and 32V pin 3 see figure 3 4 on page 3 5 c The voltage is 32V AC 3V Yes No Go to...

Страница 75: ...tep 17 14 Go to step 27 15 Perform the following a Set your VOM to read volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure to use cau...

Страница 76: ...ee figure 2 6 on page 2 27 disconnect the cable that connects test port to the siderail with the inoperative knee up function e In the cable P17 or P18 check the resistance between VCC common pin 7 an...

Страница 77: ...al injury or equipment damage a Unplug the bed from its power source SHOCK HAZARD Disconnect the battery backup Failure to do so could result in personal injury or equipment damage b If applicable dis...

Страница 78: ...e that connects the test port P200 see figure 2 6 on page 2 27 to the siderail with the inoperative knee up function at the siderail end of the cable e Check the resistance in the switch see figure 3...

Страница 79: ...tion at the siderail end of the cable e Check the resistance in the switch see figure 3 3 on page 3 4 and figure 3 4 on page 3 5 f While holding the VOM probes in place depress the knee up switch g Th...

Страница 80: ...o Go to step 19 2 Perform the following steps a Set your VOM to measure volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure to use cau...

Страница 81: ...ee figure 2 9 on page 2 36 Figure 2 9 Knee Motor Location b Set your VOM to measure volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failur...

Страница 82: ...Failure to use caution could result in serious electrical shock injury b At P1 measure the voltage between common pin 1 and 32V pin 3 see figure 3 4 on page 3 5 c The voltage is 32V AC 3V Yes No Go t...

Страница 83: ...tep 17 14 Go to step 27 15 Perform the following a Set your VOM to read volts DC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or ground Failure to use cau...

Страница 84: ...e figure 2 8 on page 2 35 disconnect the cable that connects test port to the siderail with the inoperative knee down function e In the cable P17 or P18 check the resistance between VCC common pin 7 a...

Страница 85: ...source Failure to do so could result in personal injury or equipment damage a Unplug the bed from its power source SHOCK HAZARD Disconnect the battery backup Failure to do so could result in personal...

Страница 86: ...ail with the inoperative knee down function see Siderail Nurse Call Lighting or Bed Control Switch Assembly on page 4 52 d Disconnect the cable that connects the test port P200 see figure 2 8 on page...

Страница 87: ...he test port P200 see figure 2 8 on page 2 35 to the siderail with the inoperative knee down function at the siderail end of the cable e Check the resistance in the switch see figure 3 3 on page 3 4 a...

Страница 88: ...ion now operates normally go to Final Actions on page 2 7 Otherwise go to step 5 5 Check the hilow limit switch be grasping the hilow rod and moving it back and forth approximately 3 8 cm You hear the...

Страница 89: ...cedures Page 2 44 Century Bed Service Manual man241 Figure 2 10 P200 Location c Activate the bed up switch d The voltage is 12V DC 3V Yes No Go to step 11 7 Perform the following a Disconnect the powe...

Страница 90: ...h live terminals wires or ground Failure to use caution could result in serious electrical shock injury c Using the VOM check the voltage in the power cable see figure 3 4 on page 3 5 d The voltage is...

Страница 91: ...e voltage is 32V AC 3V Yes No Go to step 14 12 Perform the following a Set your VOM to read volts AC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or groun...

Страница 92: ...live terminals wires or ground Failure to use caution could result in serious electrical shock injury b At P3 measure the voltage for the hilow motor pins 2 and 3 see figure 3 4 on page 3 5 c The vol...

Страница 93: ...and figure 3 4 on page 3 5 f The VOM ohm reading is open near infinite Yes No Go to step 26 24 Perform the following SHOCK HAZARD Unplug the bed from its power source Failure to do so could result in...

Страница 94: ...page 4 8 c Open the siderail with the inoperative bed up function see Siderail Nurse Call Lighting or Bed Control Switch Assembly on page 4 52 d Disconnect the cable that connects the test port P200 s...

Страница 95: ...4 on page 3 5 f The VOM reads open near infinite Yes No Replace the bed up function switch see Siderail Nurse Call Lighting or Bed Control Switch Assembly on page 4 52 If this solves the problem go t...

Страница 96: ...M probes in place depress the bed up switch g The VOM reads 100 or less Yes No Replace the bed up function switch see Siderail Nurse Call Lighting or Bed Control Switch Assembly on page 4 52 If this s...

Страница 97: ...ction now operates normally go to Final Actions on page 2 7 Otherwise go to step 5 5 Check the hilow limit switch be grasping the hilow rod and moving it back and forth approximately 3 8 cm You hear t...

Страница 98: ...edures Century Bed Service Manual man241 Page 2 53 2 Figure 2 12 P200 Location c Activate the bed down switch d The voltage is 12V DC 3V Yes No Go to step 11 7 Perform the following a Disconnect the p...

Страница 99: ...live terminals wires or ground Failure to use caution could result in serious electrical shock injury c Using the VOM check the voltage in the power cable see figure 3 4 on page 3 5 d The voltage is...

Страница 100: ...he voltage is 32V AC 3V Yes No Go to step 14 12 Perform the following a Set your VOM to read volts AC SHOCK HAZARD Use extreme care when checking live voltage Do not touch live terminals wires or grou...

Страница 101: ...live terminals wires or ground Failure to use caution could result in serious electrical shock injury b At P3 measure the voltage for the hilow motor pins 2 and 3 see figure 3 4 on page 3 5 c The volt...

Страница 102: ...n 2 see figure 3 1 on page 3 3 and figure 3 4 on page 3 5 f The VOM ohm reading is open near infinite Yes No Go to step 26 24 Perform the following SHOCK HAZARD Unplug the bed from its power source Fa...

Страница 103: ...equipment damage b If applicable disconnect the battery backup see Battery Backup Box on page 4 8 c Open the siderail with the inoperative bed down function see Siderail Nurse Call Lighting or Bed Co...

Страница 104: ...ts the test port P200 see figure 2 12 on page 2 53 to the siderail with the inoperative bed down function at the siderail end of the cable e Check the resistance in the switch see figure 3 3 on page 3...

Страница 105: ...ction at the siderail end of the cable e Check the resistance in the switch see figure 3 3 on page 3 4 and figure 3 4 on page 3 5 f While holding the VOM probes in place depress the bed down switch g...

Страница 106: ...to measure V DC At P1 measure the voltage between pin 1 and pin 3 see figure 3 4 on page 3 5 and figure 3 6 on page 3 6 Activate the head up switch The voltage is 12V DC 3V Yes No Replace the automat...

Страница 107: ...to measure V DC At P1 measure the voltage between pin 2 and pin 4 see figure 3 4 on page 3 5 and figure 3 6 on page 3 6 Activate the head down switch The voltage is 12V DC 3V Yes No Replace the automa...

Страница 108: ...se Trendelenburg Function Does Not Work The bed lowers completely but does not go into Trendelenburg or Reverse Trendelenburg when activated 1 Visually inspect the Trendelenburg or Reverse Trendelenbu...

Страница 109: ...p 7 5 Connect or replace the Trendelenburg or Reverse Trendelenburg linkages as necessary see Trendelenburg or Reverse Trendelenburg Assembly on page 4 36 The Trendelenburg or Reverse Trendelenburg fu...

Страница 110: ...8 cm to 10 cm and stop The hilow motor continues to run until it reaches its low limit or until the down button is released 10 Adjust or replace the Trendelenburg or Reverse Trendelenburg switches or...

Страница 111: ...before working under the bed Failure to do so could result in personal injury 2 Place bed stands under the upper lift arm pivots or 2 x 4s between the bed frame and lift arms 3 Examine the siderail st...

Страница 112: ...ap B near the slide rods C If this solves the problem go to Final Actions on page 2 7 Otherwise proceed to step 6 6 Raise the siderail to the up position and apply grease to the grease ports A see fig...

Страница 113: ...2 11 The Siderail Is Hard to Rotate Up or Down Chapter 2 Troubleshooting Procedures Page 2 68 Century Bed Service Manual man241 NOTES...

Страница 114: ...241 Page 3 1 3 Chapter 3 Theory of Operation Chapter Contents Electrical System 3 3 Theory of Operation 3 7 Electrical System 3 7 Motors 3 8 Patient Controls 3 8 Circuit Board Locations 3 10 Control B...

Страница 115: ...Chapter 3 Theory of Operation Page 3 2 Century Bed Service Manual man241 NOTES...

Страница 116: ...Electrical System Chapter 3 Theory of Operation Century Bed Service Manual man241 Page 3 3 3 Electrical System Figure 3 1 Test Port Cable Assembly Schematic m241_030...

Страница 117: ...Theory of Operation Page 3 4 Century Bed Service Manual man241 Figure 3 2 P C Board Wiring Diagram Siderail Interface Board P N 44578 Figure 3 3 SideCom Communication System Siderail Bed Control Keys...

Страница 118: ...of Operation Century Bed Service Manual man241 Page 3 5 3 Figure 3 4 Wiring Diagram Label Figure 3 5 Bed Exit II General Schematic Refer to fold out FO 3 1 at the rear of this manual Refer to fold out...

Страница 119: ...Page 3 6 Century Bed Service Manual man241 Figure 3 6 Control Board Schematic Figure 3 7 P C Board Wiring Diagram Night Light Board P N 43200 01 Refer to fold out FO 3 3 at the rear of this manual Ref...

Страница 120: ...uired to make this bed meet applicable electrical codes Each bed is factory tested for complete operation with and without load Each bed is tested for insulation integrity and micro leakage currents T...

Страница 121: ...until the thermal resets Patient Controls Patient operation of the various functions of this bed are accomplished by means of low voltage finger touch switches located in a convenient fixed position a...

Страница 122: ...flat position The flat feature is an overriding function that can be acquired even though the bed has been positioned in automatic contour As prescribed by the attending physician this bed has the cap...

Страница 123: ...Optional Functions Figure 3 9 Circuit Board Locations Symbol Circuit Board Circuit Board Location A Control board Left foot end of the bed B Siderail interface board Bottom channel of the siderails C...

Страница 124: ...hich drives limit switches siderails attendant control console and the Trendelenburg function This information will help you understand the control boards and act as an aid in troubleshooting problems...

Страница 125: ...Theory of Operation Chapter 3 Theory of Operation Page 3 12 Century Bed Service Manual man241 NOTES...

Страница 126: ...eplacement 4 6 Adjustment 4 7 Battery Backup Box 4 8 Removal 4 8 Replacement 4 10 Adjustment 4 11 Battery Backup Circuit Board 4 12 Removal 4 12 Replacement 4 14 Adjustment 4 14 Battery Backup Cells 4...

Страница 127: ...4 29 Hilow Limit Switch 4 30 Removal 4 30 Replacement 4 31 Adjustment 4 32 Automatic Contour Assembly Beds with Optional Automatic Contour Only 4 33 Removal 4 33 Replacement 4 34 Adjustment 4 35 Trend...

Страница 128: ...unication System Entertainment Control 4 49 Removal 4 49 Replacement 4 51 Adjustment 4 51 Siderail Nurse Call Lighting or Bed Control Switch Assembly 4 52 Removal 4 52 Replacement 4 54 Adjustment 4 54...

Страница 129: ...Chapter 4 Removal Replacement and Adjustment Procedures Page 4 4 Century Bed Service Manual man241 NOTES...

Страница 130: ...Failure to do so could result in personal injury or equipment damage 2 If applicable disconnect the battery backup see Battery Backup Box on page 4 8 3 If necessary remove the mattress from the bed WA...

Страница 131: ...he control board D by lifting up and pulling towards the head end of the bed Replacement 1 Insert the control board D into the control board box and secure it with the screw E 2 Connect the connectors...

Страница 132: ...backup see Battery Backup Box on page 4 8 8 Untie the foot section of the sleep surface and carefully lower it back down 9 If applicable install the mattress on the bed 10 Plug the bed into an approp...

Страница 133: ...in personal injury or equipment damage 3 Unplug the bed from its power source 4 If necessary remove the mattress from the bed WARNING When the head or foot section is folded back tie it securely to t...

Страница 134: ...4 2 Battery Backup Box Chapter 4 Removal Replacement and Adjustment Procedures Century Bed Service Manual man241 Page 4 9 4 Figure 4 2 Battery Backup Box m241_141...

Страница 135: ...ll the second screw E and lockwasher F through the tab J and into the Trendelenburg tube H 4 Route the power cords B along the right side channel K across the motor mount channel L and along the left...

Страница 136: ...4 11 4 WARNING Powered bed mechanisms can cause serious injury Operate the bed only with persons clear of mechanisms 13 Test the bed functions to ensure they are working properly NOTE Leave the bed pl...

Страница 137: ...its power source 4 If necessary remove the mattress from the bed WARNING When the head or foot section is folded back tie it securely to the bed frame before working on the bed Failure to do so could...

Страница 138: ...4 3 Battery Backup Circuit Board Chapter 4 Removal Replacement and Adjustment Procedures Century Bed Service Manual man241 Page 4 13 4 Figure 4 3 Battery Backup Circuit Board m241_144...

Страница 139: ...r cord F to the circuit board H 3 Install the battery box cover B on the battery box C and using the T25 Torx head screwdriver secure it with the screws A 4 Connect the battery backup box by plugging...

Страница 140: ...f necessary remove the mattress from the bed WARNING When the head or foot section is folded back tie it securely to the bed frame before working on the bed Failure to do so could result in personal i...

Страница 141: ...4 4 Battery Backup Cells Chapter 4 Removal Replacement and Adjustment Procedures Page 4 16 Century Bed Service Manual man241 Figure 4 4 Battery Backup Cells m241_143...

Страница 142: ...the battery box C 3 Attach the battery cables F to the battery cells D being sure to make the right connections 4 Install the battery box cover B on the battery box C and using the T25 Torx head screw...

Страница 143: ...e to do so could result in personal injury 2 Place bed stands under the upper lift arm pivots or 2 x 4s between the bed frame and lift arms 3 Using the knee up switch raise the knee section to its hig...

Страница 144: ...HAZARD Disconnect the battery backup Failure to do so could result in personal injury or equipment damage 7 If applicable disconnect the battery backup see Battery Backup Box on page 4 8 8 Unplug the...

Страница 145: ...d motor A or knee motor B with the two bolts D 3 Using the T25 Torx head screwdriver attach the ground strap with the screw E 4 Attach the connector C to the motor 5 If applicable connect the battery...

Страница 146: ...nect the battery backup Failure to do so could result in personal injury or equipment damage 2 If applicable disconnect the battery backup see Battery Backup Box on page 4 8 WARNING Place bed stands u...

Страница 147: ...eedle nose pliers remove the cotter pin D from the end of the drive screw 6 Remove the clevis pin E from the end of the head drive screw A or knee drive screw B and remove the drive screw from the bed...

Страница 148: ...pplicable see Head or Knee Motor on page 4 18 5 If applicable connect the battery backup see Battery Backup Box on page 4 8 6 Remove the bed stands or 2 x 4s from the bed frame as applicable 7 Plug th...

Страница 149: ...k tie it securely to the bed frame before working on the bed Failure to do so could result in personal injury or equipment damage 4 Lift the foot section of the sleep surface fold it back toward the h...

Страница 150: ...ocket and ratchet remove the bolts D nuts F spacers E and washers G 11 Pull the hilow motor off the end of the drive shaft and remove it from the bed Replacement 1 Push the hilow motor onto the end of...

Страница 151: ...wer connector A 5 If applicable connect the battery backup see Battery Backup Box on page 4 8 6 Plug the bed into an appropriate power source WARNING Powered bed mechanisms can cause serious injury Op...

Страница 152: ...Motor on page 4 24 SHOCK HAZARD Unplug the bed from its power source Failure to do so could result in personal injury or equipment damage SHOCK HAZARD Unplug the bed from its power source Disconnect...

Страница 153: ...nd washers D 6 Slide the hilow drive screw assembly F towards the foot end of the bed so that the end of the drive screw G comes out of the mounting bracket hole H NOTE You may have to angle the hilow...

Страница 154: ...ds B into the holes at the head end of the bed 4 Secure the hilow tie rods B with the cotter pins A 5 If applicable attach the hilow motor to the bed see Hilow Motor on page 4 24 6 If applicable conne...

Страница 155: ...result in personal injury or equipment damage 2 If applicable disconnect the battery backup see Battery Backup Box on page 4 8 3 If necessary remove the mattress from the bed WARNING When the head or...

Страница 156: ...is until all of the terminals are changed 8 Remove the two screws F and the switch bracket H being careful not to lose the spring G 9 Remove and discard the old switch E Replacement 1 Insert the new s...

Страница 157: ...the mattress on the bed 6 If applicable connect the battery backup see Battery Backup Box on page 4 8 7 Plug the bed into an appropriate power source WARNING Powered bed mechanisms can cause serious...

Страница 158: ...s Removal SHOCK HAZARD Unplug the bed from its power source Failure to do so could result in personal injury or equipment damage 1 Unplug the bed from its power source SHOCK HAZARD Disconnect the batt...

Страница 159: ...contour cover E from the automatic contour box C 7 Loosen and remove the screws F inside the automatic contour box C and separate the automatic contour assembly from the bed Replacement 1 Install the...

Страница 160: ...cable connect the battery backup see Battery Backup Box on page 4 8 6 Install the control board box cover see Control Board on page 4 5 7 Plug the bed into an appropriate power source WARNING Powered...

Страница 161: ...ft arms before working under the bed Failure to do so could result in personal injury 2 Place bed stands under the upper lift arm pivots or 2 x 4s between the bed frame and lift arms SHOCK HAZARD Unpl...

Страница 162: ...enburg or Reverse Trendelenburg Assembly Chapter 4 Removal Replacement and Adjustment Procedures Century Bed Service Manual man241 Page 4 37 4 Figure 4 11 Trendelenburg or Reverse Trendelenburg Assemb...

Страница 163: ...r how the assembly is fastened to the bed 2 Insert the Reverse Trendelenburg rod B or Trendelenburg rod C into the lever D 3 Secure the Reverse Trendelenburg rod B or Trendelenburg rod C with the cott...

Страница 164: ...ause serious injury Operate the bed only with persons clear of mechanisms 3 Lower the bed to the low position using the hilow function SHOCK HAZARD Unplug the bed from its power source Failure to do s...

Страница 165: ...before removal This will aid you during the assembly procedure 9 Pull the upper brake steer pedal through the brake block mechanism B until the cams C fall out Replacement 1 Install the cams C by pus...

Страница 166: ...Brake and Steer Chapter 4 Removal Replacement and Adjustment Procedures Century Bed Service Manual man241 Page 4 41 4 5 Plug the bed into an appropriate power source 6 Test the brakes to ensure that t...

Страница 167: ...ams see Brake Cam Assembly Beds with Central Brake and Steer on page 4 39 SHOCK HAZARD Unplug the bed from its power source Failure to do so could result in personal injury or equipment damage 2 Unplu...

Страница 168: ...will aid during assembly 6 If necessary remove the seven screws C in the bottom of the brake block mechanism B and disassemble the adapter plate and bearings Replacement 1 If applicable assemble the...

Страница 169: ...l Replacement and Adjustment Procedures Page 4 44 Century Bed Service Manual man241 5 If applicable connect the battery backup see Battery Backup Box on page 4 8 6 Plug the bed into an appropriate pow...

Страница 170: ...orx head screwdriver Combination wrench set Socket wrench set Ratchet drive 5 32 allen wrench 5 mm allen wrench Removal 1 Using the T25 Torx head screwdriver remove the screws A and separate the plast...

Страница 171: ...disconnect the battery backup see Battery Backup Box on page 4 8 6 Store pad and tie the siderails on one side of the bed WARNING Get additional personnel to help you turn the bed over on the side th...

Страница 172: ...he other half of the brake feature The other two casters are basic swivel type casters and do not require any adjustment 1 Using the T25 Torx head screwdriver remove the screws A and separate the plas...

Страница 173: ...brake 10 Turn the setscrew C down clockwise two full turns 11 Test the brake caster F to ensure that It does not move when the brake steer pedal is in the brake position It rolls freely when the brake...

Страница 174: ...ed from its power source Failure to do so could result in personal injury or equipment damage 1 Unplug the bed from its power source SHOCK HAZARD Disconnect the battery backup Failure to do so could r...

Страница 175: ...nd lift the entertainment control assembly out of the siderail 7 If applicable remove the three screws I holding the bed exit circuit board in place NOTE Note the orientation of all components before...

Страница 176: ...e a speaker insert the new speaker into the entertainment control assembly Using the 1 4 nut driver secure it with the nuts L and attach the connectors 4 If applicable install the bed exit circuit boa...

Страница 177: ...e cutters Soldering iron 60 40 lead tin rosin core solder Removal SHOCK HAZARD Unplug the bed from its power source Failure to do so could result in personal injury or equipment damage 1 Unplug the be...

Страница 178: ...the two screws F and slide the switch housing G out of the siderail 6 Locate the switch you want to replace Remove the switch from the switch housing G 7 Compare the old switch and the new switch sid...

Страница 179: ...3 16 4 8 mm c Push the ends into the splice terminal of the corresponding color wire on the new switch and crimp the terminal 10 Repeat step 8 or step 9 as applicable for a nurse call switch H or lig...

Страница 180: ...r the bed Failure to do so could result in personal injury 2 Place bed stands under the upper lift arm pivots or 2 x 4s between the bed frame and lift arms SHOCK HAZARD Unplug the bed from its power s...

Страница 181: ...wire ties D securing the cable to the bed frame 9 Using the T25 Torx head screwdriver remove the two screws C and separate the siderail from the bed Replacement 1 Place the siderail in the correct pos...

Страница 182: ...ee Battery Backup Box on page 4 8 6 Remove the bed stands or 2 x 4s from the bed frame as applicable 7 Plug the bed into an appropriate power source WARNING Powered bed mechanisms can cause serious in...

Страница 183: ...lift arms before working under the bed Failure to do so could result in personal injury 2 Place bed stands under the upper lift arm pivots or 2 x 4s between the bed frame and lift arms SHOCK HAZARD U...

Страница 184: ...le B from the control box 8 Loosen the thumbscrew D on the pendant holder C and remove the pendant control A Replacement 1 Mount the pendant control A on the pendant holder C and tighten the thumbscre...

Страница 185: ...le B to the control box 5 If applicable connect the battery backup see Battery Backup Box on page 4 8 6 Remove the bed stands or 2 x 4s from the bed frame as applicable 7 Plug the bed into an appropri...

Страница 186: ...and degrees Follow these steps to adjust the indicators WARNING Powered bed mechanisms can cause serious injury Operate the bed only with persons clear of mechanisms 1 Raise the bed to the high posit...

Страница 187: ...nburg and Reverse Trendelenburg Indicator Chapter 4 Removal Replacement and Adjustment Procedures Page 4 62 Century Bed Service Manual man241 Figure 4 19 Trendelenburg and Reverse Trendelenburg Indica...

Страница 188: ...12 Surface Module 5 15 Frame Surface Module 5 16 Hilow Drive Assembly 5 20 DC Drive Module 5 24 DC Head Drive Module with CPR Release 5 26 Battery Backup Assembly 5 28 Automatic Contour Cable Routing...

Страница 189: ...an241 SideCom Communication System Head Siderails Cable Routing 5 50 SideCom Communication System Head Siderail Assembly 5 52 SideCom Communication System Head Siderail Entertainment Controls 5 56 Sid...

Страница 190: ...number serial number and description of problem A qualified specialist will provide via telephone United States and Canada or FAX Outside the United States and Canada troubleshooting assistance for f...

Страница 191: ...Warranty Chapter 5 Parts List Page 5 4 Century Bed Service Manual man241 NOTES...

Страница 192: ...identification label A see figure 5 1 on page 5 5 Figure 5 1 Product Identification Label Location Call Hill Rom Technical Support at 800 445 3720 with the following information Six digit customer acc...

Страница 193: ...OM INC 1069 STATE ROUTE 46 E BATESVILLE IN 47006 9167 Address all return goods to ATTN SERVICE STORES DISTRIBUTION CENTER DOOR D23 HILL ROM INC COUNTY ROAD 300E BATESVILLE IN 47006 9167 NOTE To elimin...

Страница 194: ...es the invoice accompanying the parts will show the full selling price only for internal use at Hill Rom Do not confuse this price with your price Do not return any parts without an RMA number When pa...

Страница 195: ...Exchange Policy Chapter 5 Parts List Page 5 8 Century Bed Service Manual man241 NOTES...

Страница 196: ...1 amp fuse as applicable 4314312 1400 10 0 5 amp fuse 6336806 1400 2 6 amp circuit breaker 6340101 1400 2 DC motor head 6340102 1400 2 DC motor knee 6340103 1400 2 DC motor hilow 6219002 1400 2 Hilow...

Страница 197: ...10 Century Bed Service Manual man241 Base Module Figure 5 2 Base Module Table 5 2 Base Module Item Number Part Number Quantity Description 1 44299 1400 1 Base weldment 2 43878 1400 8 Torx button head...

Страница 198: ...Base Module Chapter 5 Parts List Century Bed Service Manual man241 Page 5 11 5 NOTES...

Страница 199: ...nits With Central Brake and Steer Chapter 5 Parts List Page 5 12 Century Bed Service Manual man241 Base Module Units With Central Brake and Steer Figure 5 3 Base Module Units With Central Brake and St...

Страница 200: ...ft arm assembly 11 33562 1400 2 Torque tube 12 11093 1400 2 Cotter pin 13 10902 1400 2 Washer 14 31113 1400 0 or 4 Leg cover 15 44100 1400 0 or 4 Leg cover 16 43878 1400 8 Torx button head screw 17 43...

Страница 201: ...low screw 44 42247 1400 1 Mechanism block bottom 45 18863 1400 1 Roller 46 43854 1400 1 Follower shaft 47 43844 1400 2 Bearing 48 12220 1400 2 Tru arc ring 49 34715 1400 2 Rocker arm assembly 50 11094...

Страница 202: ...vice Manual man241 Page 5 15 5 Surface Module Figure 5 4 Surface Module Table 5 4 Surface Module Item Number Part Number Quantity Description 1 45884 1400 1 Seat pan 2 43878 1400 2 Torx button head sc...

Страница 203: ...Frame Surface Module Chapter 5 Parts List Page 5 16 Century Bed Service Manual man241 Frame Surface Module Figure 5 5 Frame Surface Module m241_004...

Страница 204: ...1400 1 Thigh section assembly 13 45884 1400 1 Seat pan 14 32621 1400 1 Lower lift arm assembly 15 24449 1400 4 Retaining ring 16 8252 1400 As required Oil M 1 2 ounce 17 24452 1400 4 Bushing 18 60161...

Страница 205: ...0 2 Trendelenburg assembly 40 17291 1400 4 Fastener 41 24798 1400 2 Spring 42 27804 1400 2 Trip lever 43 SA1372 1400 2 Trendelenburg mount and retainer 44 3844 1400 2 Washer 45 451 1400 2 Screw 46 SA3...

Страница 206: ...Frame Surface Module Chapter 5 Parts List Century Bed Service Manual man241 Page 5 19 5 NOTES...

Страница 207: ...Hilow Drive Assembly Chapter 5 Parts List Page 5 20 Century Bed Service Manual man241 Hilow Drive Assembly Figure 5 6 Hilow Drive Assembly m241_001...

Страница 208: ...k 10 124 1400 2 Roll pin 11 831 1400 2 Locknut 12 35667 1400 2 Washer 13 39812 1400 1 Brake bracket 14 39315 1400 1 Retaining ring 15 755 1400 2 Locknut 16 39811 1400 1 Nut bracket 17 39315 1400 1 Ret...

Страница 209: ...33030 1400 1 Hilow limit rod 33 39652 1400 1 Bushing 34 43878 1400 1 Torx button head screw 35 42889 1400 2 Hex bolt 36 28082 1400 1 Brake spring 37 40741 1400 2 Thrust race 38 36250 1400 1 Coupling a...

Страница 210: ...Hilow Drive Assembly Chapter 5 Parts List Century Bed Service Manual man241 Page 5 23 5 NOTES...

Страница 211: ...DC Drive Module Chapter 5 Parts List Page 5 24 Century Bed Service Manual man241 DC Drive Module Figure 5 7 DC Drive Module m241_003...

Страница 212: ...button head screw 4 6340803 1400 1 Knee drive assembly without CPR 5 10902 1400 2 Washer 6 11093 1400 2 Cotter pin 7 17405 1400 2 Cotter pin 8 6340901 1400 0 or 1 Hilow drive 80 bed standard 9 634090...

Страница 213: ...DC Head Drive Module with CPR Release Chapter 5 Parts List Page 5 26 Century Bed Service Manual man241 DC Head Drive Module with CPR Release Figure 5 8 DC Head Drive Module with CPR Release m241_002...

Страница 214: ...cable CPR 2 22322 1400 2 Lock washer 3 35112 1400 2 Flat washer 4 43090 1400 2 Shoulder bolt hex 5 20802 1400 6 KEPS nut 6 49287 1400 1 Head screw mounting bracket 7 49285 1400 1 Head screw strap clam...

Страница 215: ...Battery Backup Assembly Chapter 5 Parts List Page 5 28 Century Bed Service Manual man241 Battery Backup Assembly Figure 5 9 Battery Backup Assembly m241_142...

Страница 216: ...7 6381402 1400 0 or 1 PCB assembly 230V 8 6469803 1400 1 Serial number label 9 64519 1400 1 Battery box lid weldment 10 43878 1400 2 Torx button head screw 11 6469802 1400 1 Warning label 12 64554 14...

Страница 217: ...Automatic Contour Cable Routing Chapter 5 Parts List Page 5 30 Century Bed Service Manual man241 Automatic Contour Cable Routing Figure 5 10 Automatic Contour Cable Routing m241_006...

Страница 218: ...ew 6 43878 1400 13 Torx button head screw 7 19124 1400 5 Large cable tie 8 40497 1400 5 KEPS nut 9 23208 1400 2 Washer 10 6342201 1400 0 or 1 Power cord 110V domestic 11 6342202 1400 0 or 1 Power cord...

Страница 219: ...Automatic Contour Module Chapter 5 Parts List Page 5 32 Century Bed Service Manual man241 Automatic Contour Module Figure 5 11 Automatic Contour Module m241_005...

Страница 220: ...43878 1400 13 Torx button head screw 13 34422 1400 1 Cable retainer 14 14451 1400 4 Cable clamp 15 3302002 1400 0 or 1 Automatic contour assembly 16 63505 1400 1 Power cable low voltage 17 25647 1400...

Страница 221: ...1400 1 Small cable tie 35 9023506 1400 1 Screw 36 31236 1400 1 Lockwasher 37 63485 1400 1 Motor harness 38 3292101 1400 1 Control box cover 39 35186 1400 1 Sound dampening foam 40 63421 1400 1 Wiring...

Страница 222: ...Automatic Contour Module Chapter 5 Parts List Century Bed Service Manual man241 Page 5 35 5 NOTES...

Страница 223: ...Head Drive Assembly P N 6340802 Chapter 5 Parts List Page 5 36 Century Bed Service Manual man241 Head Drive Assembly P N 6340802 Figure 5 12 Head Drive Assembly P N 6340802 m241_069...

Страница 224: ...3 1400 2 Bearing 7 11578 1400 1 Thrust bearing 8 43878 1400 4 Torx button head screw 9 10640 1400 1 Roll pin 10 63477 1400 2 Head screw cover 11 27717 1400 2 Clip 12 32638 1400 1 Roller 13 34680 1400...

Страница 225: ...mbly with CPR Module P N 6340804 Chapter 5 Parts List Page 5 38 Century Bed Service Manual man241 Head Drive Assembly with CPR Module P N 6340804 Figure 5 13 Head Drive Assembly with CPR Module P N 63...

Страница 226: ...Screw 9 43878 1400 4 Torx button head screw 10 63492 1400 2 Screw cover 11 34680 1400 4 Screw 12 38429 1400 1 Head limit rod 13 38122 1400 1 Bearing housing 14 33615 1400 2 Spacer 15 33621 1400 2 Head...

Страница 227: ...11579 1400 2 Thrust washer 43 45713 1400 1 CPR release cable assembly 44 33279 1400 1 Head screw bracket 45 39237 1400 1 Head flat ball screw 46 40672 1400 1 Stop head screw 47 33622 1400 1 Bearing sp...

Страница 228: ...dule P N 6340804 Chapter 5 Parts List Century Bed Service Manual man241 Page 5 41 5 64 32589 1400 2 Mounting plate 65 36250 1400 1 Coupling assembly 66 63464 1400 1 Motor mount weldment Item Number Pa...

Страница 229: ...Knee Drive Assembly P N 6340803 Chapter 5 Parts List Page 5 42 Century Bed Service Manual man241 Knee Drive Assembly P N 6340803 Figure 5 14 Knee Drive Assembly P N 6340803 m241_068...

Страница 230: ...0 2 Mounting plate 6 34613 1400 2 Bearing 7 32596 1400 1 Spacer 8 43878 1400 4 Torx button head screw 9 63491 1400 2 Knee screw cover 10 10640 1400 1 Roll pin 11 27717 1400 2 Clip 12 34680 1400 8 Scre...

Страница 231: ...Night Light Assembly P N 4578201 Chapter 5 Parts List Page 5 44 Century Bed Service Manual man241 Night Light Assembly P N 4578201 Figure 5 15 Night Light Assembly P N 4578201 m241_130...

Страница 232: ...in relief 11 28837 1400 2 Hex nut 12 29975 1400 2 Pin receptacle 13 3091002 1400 1 Plug housing 14 9000201 1400 0 66 lft Tubing blank 15 3091301 1400 1 Plug housing socket 3 position 16 30950 1400 4 T...

Страница 233: ...Pendant Control Assembly P729 Chapter 5 Parts List Page 5 46 Century Bed Service Manual man241 Pendant Control Assembly P729 Figure 5 16 Pendant Control Assembly P729 m241_131...

Страница 234: ...0 1 Thumbscrew 3 SA4516 1400 1 Pendant sleeve 4 SA7089 1400 1 Rubber grommet 5 SA7088 1400 1 Pendant holder 6 SA7091 1400 1 C ring 7 43615 1400 1 Velcro strap 8 19124 1400 1 Large cable tie 9 SA7346 1...

Страница 235: ...l man241 Head Siderail Mounting Parts Figure 5 17 Head Siderail Mounting Parts Table 5 17 Head Siderail Mounting Parts Item Number Part Number Quantity Description 1 43878 1400 10 Torx button head scr...

Страница 236: ...49 5 Dust Cover Module Figure 5 18 Dust Cover Module Table 5 18 Dust Cover Module Item Number Part Number Quantity Description 1 44489 1400 12 Six lobe pan head screw 2 32967 1400 0 or 1 Head end cove...

Страница 237: ...ead Siderails Cable Routing Chapter 5 Parts List Page 5 50 Century Bed Service Manual man241 SideCom Communication System Head Siderails Cable Routing Figure 5 19 SideCom Communication System Head Sid...

Страница 238: ...0 2 Slide bracket screws 4 34778 1400 1 Caution label 5 21341 1400 2 Cord holder 6 29788 1400 4 Twist nut 7 4759 1400 5 Screw 8 20312 1400 3 Cable clamp 9 33358 1400 8 Plug button 10 42006 1400 2 Scre...

Страница 239: ...System Head Siderail Assembly Chapter 5 Parts List Page 5 52 Century Bed Service Manual man241 SideCom Communication System Head Siderail Assembly Figure 5 20 SideCom Communication System Head Sidera...

Страница 240: ...communication cable assembly rh 12 4455503 1400 1 Siderail communication cable assembly lh 13 28717 1400 4 Bushing 14 48887 1400 2 Siderail insert female 15 49223 1400 2 Locking plate 16 9018808 1400...

Страница 241: ...ack rh 46 4446501 1400 1 Communication housing back lh 47 31273 1400 0 or 2 Nurse call insert blank 48 31274 1400 0 or 2 Nurse call insert blank 49 31652 1400 1 Control housing rh 50 31653 1400 1 Cont...

Страница 242: ...SideCom Communication System Head Siderail Assembly Chapter 5 Parts List Century Bed Service Manual man241 Page 5 55 5 NOTES...

Страница 243: ...Entertainment Controls Chapter 5 Parts List Page 5 56 Century Bed Service Manual man241 SideCom Communication System Head Siderail Entertainment Controls Figure 5 21 SideCom Communication System Head...

Страница 244: ...0 As required Solder 7 35290 1400 2 Volume control switch 8 35171 1400 6 Shrink tubing 9 63496 1400 2 Label switch overlay 10 63494 1400 2 Label SideCom Communication Sys tem switches 11 28286 1400 4...

Страница 245: ...m Head Siderail Bed Controls Chapter 5 Parts List Page 5 58 Century Bed Service Manual man241 SideCom Communication System Head Siderail Bed Controls Figure 5 22 SideCom Communication System Head Side...

Страница 246: ...abel SideCom Communication Sys tem switches 5 31659 1400 1 Nurse call insert rh 6 31660 1400 1 Nurse call insert lh 7 63496 1400 2 Label switch overlay 8 31661 1400 2 Light insert 9 28493 1400 2 LED y...

Страница 247: ...SideCom Communication System Head Siderail Bed Controls Chapter 5 Parts List Page 5 60 Century Bed Service Manual man241 NOTES...

Страница 248: ...Hard to Clean Spots 6 3 Disinfection 6 3 Component Handling 6 4 P C Boards 6 4 Lubrication Requirements 6 5 Preventive Maintenance 6 6 Preventive Maintenance Schedule 6 7 Preventive Maintenance Check...

Страница 249: ...Chapter 6 General Procedures Page 6 2 Century Bed Service Manual man241 NOTES...

Страница 250: ...mage could occur CAUTION Do not use harsh cleaners solvents or detergents Equipment damage could occur General Cleaning Clean the unit with a lightly dampened cloth and ordinary disinfectants Do not u...

Страница 251: ...can cause damage to the P C boards shorten their service life and result in bed malfunctions The following rules for handling should always be observed Wear an appropriate antistatic electrostatic di...

Страница 252: ...the system By retaining oil the pores give a self lubricating quality to the bearings and bushings If any silicone based lubricant is applied to the bearings and bushings or anywhere else on the syste...

Страница 253: ...will help ensure a long operative life for the Century Bed PM will minimize downtime due to excessive wear The following PM schedule guides the technician through a normal PM procedure on the Century...

Страница 254: ...re the bed does not move when the brakes are activated Adjust the brakes if necessary Inspect and adjust the steering mechanism if necessary Caster tires Check the caster tires for cuts wear tread etc...

Страница 255: ...imits Automatic contour limits Trendelenburg Drive screws Lockout switches Brake and steer Caster tires Siderail controls Model Number Siderail frame Pivot points Power cord and plug Head and foot pan...

Страница 256: ...ead screwdriver 1 4 nut driver Ratchet drive Socket wrench set 1 2 socket and ratchet 7 16 socket and ratchet 1 2 open end wrench Combination wrench set Wire cutters Needle nose pliers Channellock pli...

Страница 257: ...he Mod I B and Mod IV C receptacle connectors see figure 6 1 on page 6 10 Figure 6 1 Extraction Tool P N 91065 1 The extraction tool tip D is designed to enter the lance recess in the front of the hou...

Страница 258: ...ates the contacts The lance recess is the cavity located between the contact and the connector housing A solid rib locking edge at the rear of the lance recess cavity provides a stop for the extractio...

Страница 259: ...ront of the tool fits over the pin and socket contacts When the tip is bottomed in the contact cavity the contact locking lances are forced together allowing the contact to slide free from the housing...

Страница 260: ...l Procedures Century Bed Service Manual man241 Page 6 13 6 6 3 Siderail Lubrication 1 Raise the siderail to the fully upright position 2 Apply grease to the grease ports A see figure 6 2 on page 6 13...

Страница 261: ...6 3 Siderail Lubrication Chapter 6 General Procedures Page 6 14 Century Bed Service Manual man241 NOTES...

Страница 262: ...7 4 Removal 7 4 Adjustment 7 5 Architectural Bumpers P819 7 6 Installation 7 6 Removal 7 6 Adjustment 7 6 IV Rod P2217 7 7 Installation 7 7 Removal 7 7 Adjustment 7 8 Trapeze Support P846 or P846D 7...

Страница 263: ...Chapter 7 Accessories Page 7 2 Century Bed Service Manual man241 NOTES...

Страница 264: ...er bumpers P819 Architectural bumper assembly P4039HE International end panel P2217 IV rod P846 Trapeze support P846D Trapeze support Canada P9912A Bed extender P721E Four function pendant control P44...

Страница 265: ...an be equipped retrofitted with wall protecting roller bumpers Installation Using the crescent wrench install the two bolts A to secure the roller bumper B to the bed frame C see figure 7 1 on page 7...

Страница 266: ...7 1 Roller Bumpers P818C08 Chapter 7 Accessories Century Bed Service Manual man241 Page 7 5 7 Adjustment None...

Страница 267: ...l protection features The Century Bed can be equipped with wall protecting architectural bumpers Installation Using the crescent wrench install the two bolts A and secure the architectural bumper B to...

Страница 268: ...the bed two at the head end two in the seat section and two at the foot end The height of the IV rod is easily adjustable Installation Insert the IV rod into the desired socket Twist the lower section...

Страница 269: ...Century Bed Service Manual man241 Adjustment 1 To extend the IV rod pull upward on the upper section A to the desired height 2 To lower the IV rod pull outward on the release knob B and manually lowe...

Страница 270: ...ation 1 Remove the hole plugs from the bed frame at the end of bed where the trapeze support is to be installed 2 Position the trapeze support A and backing plates D on the bed frame B see figure 7 4...

Страница 271: ...e the trapeze support assembly is attached to the bed Use the control box lockout to deactivate the hilow function Equipment damage could occur Removal 1 Loosen and remove the bolts C and locknuts and...

Страница 272: ...kets have a 3 4 inside diameter while the P847C sockets have a 1 2 inside diameter Please specify which size is needed when ordering Installation 1 Insert the short adapter tubes A into the IV sockets...

Страница 273: ...Frame Adapter P847B and P847C Chapter 7 Accessories Page 7 12 Century Bed Service Manual man241 2 Remove the long adapter tubes B from the IV sockets at the foot end of the bed by lifting upward Adju...

Страница 274: ...Schematic Wiring Diagram Wiring Diagram Label Back to Chapter 3...

Страница 275: ...Schematic Wiring Diagram Bed Exit II General Schematic Back to Chapter 3...

Страница 276: ...Schematic Wiring Diagram Control Board Back to Chapter 3...

Страница 277: ...Schematic Wiring Diagram Night Light P C Board Wiring Diagram Back to Chapter 3...