TP-R-400
Operation
23
2.2.3
Inserting/Changing the Battery
NOTICE
Risk of damage to the equipment!
•
Clean and dry the probe well before opening!
•
Do not use compressed air to clean the touch probe!
•
Replace empty batteries immediately!
1. Turn battery cover lock anti-clockwise and remove the battery cover to the outside of the housing of the touch
probe.
2. Remove the used batteries.
NOTICE
Risk of damage to the equipment!
When closing the battery compartment, ensure the O-ring is properly seated and in faultless condition!
INFORMATION
The Touch Probe TP-R-400 is protected against wrong polarity. There is no need to pay attention to the polarity
when inserting new batteries, but the batteries must always be inserted alternately into the battery compartment!
3. Insert new batteries
alternately
into the battery compartment (refer to Fig. 10) and insert the battery cover into
the housing.
4. Turn the battery cover lock clockwise to lock the battery compartment.
After insertion or replacement of the battery, the touch probe initialises for approx. 5 s. The LEDs illuminate
in blue during this time.
Replacing batteries
Battery cover
(40.02-BATTERIEDECKEL)
with O-ring 20x1 for battery cover
(5186)
2x Battery (3.6 V / ½ AA)
(4316)
= Close
= Open
Assembly key for battery cover
(5840)