Operation
TP-R-400
20
2.2.1.2
Installing/Changing the Stylus (Measuring Unit PP41.00/HPP41.10)
NOTICE
Risk of damage to the equipment!
The probe mechanic is sensitive, therefore never use excessive force when screwing in the stylus!
•
The screwing torque is M=2 Nm!
•
When using a break shaft adapter (91.00-S-M4/M4), only screw in the stylus with a maximal torque of 1 Nm
(0.74 lbf.ft)!
•
The use of a break shaft adapter (91.00-S-M4/M4) is not permitted in conjunction with precision measuring
unit HPP41.10!
1. Unscrew the stylus from the touch probe using the mounting pin.
2. Carefully screw the new stylus with mounting pin into the touch probe (Fig. 7).
3. Align the stylus to spindle center (refer to chapter 2.2.5).
4. Calibrate the probe.
Stylus change with measuring unit PP41.00/HPP41.10 (using PP41.00 as an example)
Mounting pin
(0885)
Stylus
M=2 Nm (1.47 lbf.ft)
Содержание TP-R-400
Страница 1: ......