81/97
10.3 Hook in the front panel at the support sheet for electronics (A1)
In order to perform electrical measurements at the centrifuge it is necessary to
hook in the front panel at the support sheet for electronics (A1).
•
Remove the front panel and the support sheet for electronics (A1) as described in
chapters 10.1, 10.2, pg. 80.
•
Plug the flat ribbon cable on the operating panel (A2) again.
•
Hook in the front panel with the slot on the two supporting devices (pg. 80, Fig. 10,
a) at the support sheet for electronics (A1), see Fig. 11.
Fig. 11
10.4
Removing the upper part of the centrifuge housing
For removing the upper part of the centrifuge housing 2 persons are needed.
•
Remove the front panel and the support sheet for electronics (A1) as described in
chapters 10.1, 10.2, pg. 80.
•
Pull both plugs S703 and S707 of the lid lock on the electronics (A1), see pg. 93,
chapter 11.3.
•
Remove all required fixations of the cables.
•
Unscrew the fastening screws of the upper part of the centrifuge housing.
4 screws (pg. 77, Fig. 2, b ; pg. 78, Fig. 4, b) each on the left and right side of the
upper part of the housing, 6 screws on its rear. In centrifuges with cooling, unscrew
the two screws (pg. 78, Fig. 4, c) on the front of the upper part of the centrifuge
housing.
•
Remove the upper part of the centrifuge housing and place it next to the centrifuge.
•
To mount the upper part of the housing, carry out these steps in opposite order.
Содержание ROTINA 420
Страница 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Страница 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Страница 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Страница 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Страница 96: ...95 97 ...
Страница 97: ...96 97 ...