59/97
Action
Display / Comment
4. Press the
T/°C
key so
often until the
following will be
displayed:
→
e.g.
Imb Xa= 0 Ya= 1 Xm= 1 Ym= 3
Xa = actual X-value in per cent of the imbalance
sensor.
Ya = actual Y-value in per cent of the imbalance
sensor.
Xm = maximum X-value in per cent of the imbalance
sensor.
Ym = maximum Y-value in per cent of the imbalance
sensor.
5. To exit the inquiry
press the
STOP / OPEN
key.
→
e.g.
6 20 9
9 4000 5:20
Display of the centrifugation data.
6. Press the
STOP / OPEN
key.
The centrifugation run is terminated.
6.15.3
Setting the imbalance switch-off
A change of the imbalance switch-off affects all rotors.
The imbalance switch-off is specified by the indication of the difference in weight of
opposite rotor positions.
In the factory the imbalance switch-off will be adjusted with rotor 4723.
With swing-out rotors all rotor positions must be lined with
identical
hangers.
By a test run with empty rotor, and that for the rotor specified switch-off weight in one
rotor place the imbalance switch-off is checked, see following table.
With the switch-off weight specified for the rotor the drive must absolutely
switch off during the run-up.
Switch-off weights of the different rotors:
Rotor Switch-off
weight
4758 25
g
4754 25
g
4753 25
g
4723 30
g
4760 6
g
4790 6
g
4794 15
g
4795 15
g
Содержание ROTINA 420
Страница 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Страница 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Страница 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Страница 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Страница 96: ...95 97 ...
Страница 97: ...96 97 ...