49/97
6.12 Set the centrifuge model
The centrifuge model, which is displayed after switch-on can be set.
Model Description
ROTINA 420
Centrifuge without cooling
ROTINA 420 R
Centrifuge with cooling
ROTINA 420 RH
Centrifuge with option heating/cooling
The setting of the centrifuge model is only possible if the rotor is at standstill.
It is at any time possible to terminate the procedure by pressing the
STOP / OPEN
key. In this case the adjustments are not saved.
Action
Display / Comment
1. Keep the
START
key
and the
STOP / OPEN
key pressed and
switch on the mains
switch at the same
time.
→
∗∗∗
Select Menu
∗∗∗
2. Press the
PROG
key.
→
-
>
Version
3. Press the
START
key.
→
e.g.
ROTINA 420
Centrifuge model
•
Turn the control
knob .
→
e.g.
ROTINA 420 R
Set the Centrifuge model.
4. Press the
START
key.
→
→
Store Version ...
The setting will be stored.
e.g.
ROTINA 420 R
5. To exit the menu
press the
STOP / OPEN
key three times.
→
→
→
-
>
Version
∗∗∗
Select Menu
∗∗∗
e.g.
6 20 9
9 4000 5:20
Display of the centrifugation data.
Содержание ROTINA 420
Страница 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Страница 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Страница 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Страница 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Страница 96: ...95 97 ...
Страница 97: ...96 97 ...