62/97
6.17 Function
test
The individual components of the centrifuge can be tested on its function.
The function test is only possible if the rotor is at standstill
In order to be able to check the function of the two speed sensors the lid must
be opened before starting the function test.
In the menu “Test” it is possible to jump back to the parameter indicated
before by pressing the
T/°C
key.
It is at any time possible to terminate the procedure by pressing the
STOP / OPEN
key.
Action
Display / Comment
1. Keep the
START
key
and the
STOP / OPEN
key pressed and
switch on the mains
switch at the same
time.
→
∗∗∗
Select Menu
∗∗∗
2. Press the
PROG
key
so often until the
following will be
displayed:
→
-
>
Test
3. Press the
START
key.
→
e.g.
00 Software Version = 01.01
Program version
4. Press the
PROG
key.
→
01 Display Full
<
-
>
Blank
•
Turn the control
knob one step to
the left.
→
▌▌▌▌▌▌▌▌▌▌▌▌▌▌▌▌
Function test of the LC-display. All segments of the
display light up.
•
Turn the control
knob one step to
the right.
→
All segments of the display deactivated (no display)
5. Press the
PROG
key.
→
e.g.
02 Imbal act = 1 : 0 max = 3 : 4
imbalance values.
act = actual X- and Y-value in per cent of the
imbalance
sensor.
max = actual X- and Y-value in per cent of the
imbalance
sensor.
Function test of the imbalance sensor:
Shake the motor by hand. The current X- and Y-
values must changed.
Содержание ROTINA 420
Страница 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Страница 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Страница 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Страница 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Страница 96: ...95 97 ...
Страница 97: ...96 97 ...