31/97
SENSOR - ERROR 91 - 92
Imbalance sensor error.
SENSOR - ERROR 91 = x-axis, SENSOR - ERROR 92 = y-axis
The rotor slows down braked until it stops.
No further user operation possible.
•
Reset error code:
Perform a MAINS RESET.
•
Check the plug S701 and the wires.
•
Imbalance sensor (B2.2) is defective.
Measure supply voltage on plug S701 / Electronics (A1)
pin 6 – pin 1 (+5 VDC).
Measure on Electronics (A1) / plug S701, pin 6 – pin 2 and
pin 6 – pin 3 (square wave signal 5 V /100 Hz).
•
Check the function of the imbalance sensor, see pg. 62, chapter
6.17, item 5.
•
Electronics (A1) is defective.
SENSOR - ERROR 93
The temperature of the imbalance sensor is out of the valid range.
The rotor slows down braked until it stops.
No further user operation possible.
•
Reset error code:
Perform a MAINS RESET.
•
Imbalance sensor (B2.2) is defective.
Measure supply voltage on plug S701 / Electronics (A1)
pin 6 – pin 1 (+5 VDC).
Measure voltage "Tout" on plug S701 / Electronics (A1)
pin 6 – pin 4.
Valid voltage range: 0.5 V up to 2.00 V.
Voltage at 25°C / 77°F = 1.25 V (
±
40 mV).
A temperature change of 1°K causes a voltage change of 5 mV.
•
Electronics (A1) is defective.