6.14 Cooling
parameters
6.14.1
Set the temperature unit
The temperature can be entered in degrees Celsius (°C) or in degrees Fahrenheit (°F).
The setting of the temperature unit is only possible if the rotor is at standstill.
In the menu “Settings” it is possible to jump back to the parameter indicated
before by pressing the
T/°C
key.
It is at any time possible to terminate the procedure by pressing the
STOP / OPEN
key. In this case the adjustments are not saved.
Action
Display / Comment
1. Switch on the mains
switch.
→
→
→
→
e.g.
ROTINA 420R
Centrifuge model
e.g.
V 01.01
Program version
e.g.
Rotor 4 Nmax=4500 R=184
Rotor information
e.g.
6 20 9
9 4000 5:20
Display of the centrifugation data.
2. Keep the
PROG
key
pressed for
8 seconds.
Display after
8 seconds.
→
→
e.g.
RCL 4
∗∗∗
Machine Menu
∗∗∗
3. Press the
PROG
key
so often until the
following will be
displayed:
→
-
>
Settings
4. Press the
START
key.
→
Sound / Bell = off/on
Acoustic signal.
5. Press the
PROG
key
so often until the
following will be
displayed:
→
Temp Unit = Celsius/Fahrenheit
The set temperature unit.
•
Turn the control
knob .
→
Temp Unit = Celsius/Fahrenheit
Set Celsius (°C) or Fahrenheit (°F).
Содержание ROTINA 420
Страница 1: ...06 08 Andreas Hettich GmbH Co KG AR4701EN ROTINA 420 ROTINA 420 R Repair instructions ...
Страница 78: ...77 97 Fig 2 Fig 3 3 1 5 4 6 7 a b d 13 16 18 17 19 21 22 30 25 23 38 21 20 b a c d ...
Страница 80: ...79 97 Fig 6 Fig 7 Fig 8 9 10 12 11 7 8 a b 29 28 27 26 41 43 44 41 39 42 40 42 45 a c b d d ...
Страница 95: ...94 97 11 3 2 Connecting diagram ROTINA 420 420 R ...
Страница 96: ...95 97 ...
Страница 97: ...96 97 ...