EN
AB1203EN_SA
Rev. 02 / 11.2022
29/38
20.6
Querying the hours of operation and the number of centrifugation runs
The hours of operation are divided into internal and external hours of operation.
Internal hours of operation: Total time the device was switched on.
External hours of operation: Total time of the previous centrifugation runs.
With the rotor at a standstill, the query can proceed as follows:
The procedure can be aborted at any time by pressing the
STOP
OPEN
key.
SELECT
Press and hold the button for eight
seconds.
t/min:s
RPM
>RCF<
SELECT
Press the key until the following is
displayed.
t/min:s
RPM
>RCF<
START
PULSE
Press the key.
TimeExt :
External hours of operation.
Example:
t/min:s
RPM
>RCF<
SELECT
Press the key.
TimeInt :
Internal hours of operation.
Example:
t/min:s
RPM
>RCF<
SELECT
Press the key.
Starts :
Number of all centrifugation runs.
Example:
t/min:s
RPM
>RCF<
STOP
OPEN
Press the key twice to exit the "
Time &
Cycles"
menu or press three times to exit the "
MACHINE MENU ".
t/min:s
RPM
>RCF<
Example:
t/min:s
RPM
>RCF<