EN
AB1203EN_SA
Rev. 02 / 11.2022
17/38
16.2
Input of the centrifugation parameters with the "SELECT" key
The runtime can be set in minutes and seconds (parameters
t/min
and
t/sec
).
In order to set the continuous run, the parameters
t/min
and
t/sec
must be set to zero. In the time display
(t/min:s), "
--:--
" appears.
Example:
t/min:s
RPM
>RCF<
The relative centrifugal acceleration (RCF) depends on the centrifuging radius (RAD).
During the input of the RCF, the set centrifuging radius is displayed.
If no key is pressed for 8 seconds after selection or during parameter input, the previous values are shown on
the display. The parameters must then be input again.
By pressing the
START
PULSE
button, the settings will be saved.
If several parameters are input, the
START
PULSE
key must be pressed after setting the last parameter.
Entering parameters can be cancelled at any time by pressing the
STOP
OPEN
key. In this case, the settings are not
stored.
Example:
RPM display (RPM)
Example:
RCF display ( RCF )
RCF
Press the key to activate the RPM display
(RPM) or the RCF display ( RCF ) as
needed.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
SELECT
Press the key.
t/min :
Runtime, minutes.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t
Set the desired value with the keys.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
SELECT
Press the key.
t/sec :
Runtime, seconds.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t
Set the desired value with the keys.
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
Continued on next page