EN
22/38
Rev. 02 / 11.2022
AB1203EN_SA
19 Cycle
counter
The period of use of the rotor is limited to 50000 running cycles (centrifugation runs).
The centrifuge is equipped with a cycle counter, which counts the running cycles (centrifugation runs).
After each centrifugation run, while the lid is unlocking, the remaining number of running cycles (centrifugation runs)
will be briefly displayed.
Example:
t/min:s
RPM
>RCF<
If the maximum permissible number of rotor running cycles has been exceeded, the following is displayed every time
the centrifugation run is started and the centrifugation run must be restarted.
t/min:s
RPM
>RCF<
The following is displayed:
t/min:s
RPM
>RCF<
the rotor must be immediately exchanged for a new rotor for safety reasons.
After the rotor has been exchanged, the cycle counter must be reset to “0” (see chapter “Resetting the cycle counter
to ‘0’ “).
20 Settings
and
queries
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
t/min:s
RPM
>RCF<
Query:
System information
Setting:
Acoustic signal
Automatic unlocking of
the lid after the
centrifugation run
Automatic switch-off of
the background lighting
Optical signal after
ending the
centrifugation run
Query:
Hours of operation
Number of completed
centrifugation runs
Setting:
Cycle counter