Operating Manual
Page 33
8. Troubleshooting
Only qualified personal shall be carrying out repair and maintenance work.
Always wear required personal protection equipment.
Opening covers or removing parts may expose live (electrical) connections.
8.1. Display Blank After Turning ON Power
POSSIBLE CAUSE
REPAIR
Blown Fuse
Locate cause and perform necessary correctiveactions.
Then Replace Fuse.
No Supply Power
Measure input power and trace tosource.
Ribbon Cable Connection Loose Remove and reseat connector.
Contrast of LCD display too
dark or light
Adjust contrast potentiometer on processor board located on
rear of lid.
Defective Power Supply Unit
Return entire unit to Henkel for repair.
Processor PC-board defective
8.2. Configuration Error After Turning ON Power
POSSIBLE CAUSE
REPAIR
Existing modules and
configuration do not match
Check configuration and modules. Press to reconfigure.
8.3. Display Value Remains Zero Despite Altered Conductivity Value
POSSIBLE CAUSE
REPAIR
Conductivity cable connection
to sensor not correct or broken
Inspect cables and establish connection.
8.4. Display Value Unstable
POSSIBLE CAUSE
REPAIR
Conductivity – air in water
Move sensor to location without air bubbles. Position
sensor so that it doesn’t trap air.
pH/ORP – Defective Sensor or
Cables
Replace sensor or cables
All Sensors – faulty shield
connection
Replace cable or improve shield connection.
8.5. Error Message “Plausibility Error” During Calibration.
POSSIBLE CAUSE
REPAIR
Wrong calibration solutions
Use correct buffer or standard solutions.
Measurement system too
sluggish – times out
See error description “Display Too Sluggish”