29
Technical Specifications
Dimensions: 160 x 210 x 90 mm (6.29 x 8.26 x 3.54 inches)
Weight: 1,000 g (2.20 pounds)
5 alkaline type AA batteries
Power adapter: CE marked
Only use adapters, as listed under ’Adapters’.
Pollution degree: 2
Overvoltage category: II
Atmospheric pressure: 800 hPa to 1060 hPa.
Equipment not suitable for use in the presence of flammable
mixtures.
The instrument is tested according to IEC 61010-1, Second edition:
2001 and EN 61010-1: 2001, IEC/EN 61010-2-101: 2002, IEC 61326-1:
2005 and EN 61326-1: 2006, IEC 61326-2-6,
First edition: 2005 and EN 61326-2-6: 2006, IEC 60601-1-2, Third
edition: 2007 and EN 60601-1-2: 2007 and complies with the IVD
Medical Device Directive 98/79/EC.
The instrument is made for continuos mode.
Essential Performance
The essential performance is quantitative determination of
hemoglobin in plasma and serum specimens, aquenous solutions, or
stored banked erythrocytes.
AC adapters
Country: EU/US/GB
Type: 12 V Friwo FW7333SM/12 V HCA01
Input: 100 V~-240 V~/50–60 Hz/
≤
200 mA
Recommended separation distance between Portable
and mobile RF communications equipment and
HemoCue
®
Plasma/Low Hb Photometer
The HemoCue systems are intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled.
The customer or the user of HemoCue systems can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and RF communications equipment (transmitters)
and HemoCue systems as recommended below, according to the
maximum output power of the communications equipment.
Rated maxi-
mum output
power of
transmitter
(W)
Separation distance according to frequency
of transmitter (m)
150 kHz
to 80 MHz
d = 1.2
√
P
80 MHz to
800 MHz
d = 1.2
√
P
800 MHz to
2.5 GHz
d = 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at maximum output power not listed above,
the recommended separation distances (d) in meters (m) can be esti-
mated using the equation applicable to the frequency of the transmit-
ter, where (P) is the maximum output power rating of the transmitter
in watts (W) according to transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the
higher frequencyrange applies.
NOTE 2
These guidelines may not apply in all situations. Elec-
tromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Содержание Plasma/Low Hb
Страница 1: ...HemoCue Plasma Low Hb Operating manual Manual de funcionamiento Manuel d utilisation ...
Страница 2: ......
Страница 6: ...1 5 4 3 2 6 6 Components Componentes Composants ...
Страница 8: ...1 2 3a 5 3c 4 3b 8 Start up Activación Démarrage ...
Страница 10: ...5 3 4 2 2 1 10 Measuring Sample material Medición Tipo de muestra Mesure Échantillons ...
Страница 12: ...7 8 6 9 12 ...
Страница 14: ...2 1 14 Maintenance Mantenimiento Maintenance ...