28
Correlation study
1) Aqueous samples from surgical procedures, analyzed on the
HemoCue Plasma/Low Hb system with the HiCN-(ICSH) method
as reference.
2) Erythrocyte suspensions, analyzed on the HemoCue Plasma/Low
Hb system with the HiCN-(ICSH) method as reference.
3) Erythrocyte suspensions, analyzed on the HemoCue Plasma/Low
Hb system with the HiCN-(ICSH) method as reference.
Reference
HemoCue
Group
N
Mean
(g/L)
Min
(g/L)
Max
(g/L)
SD
(g/L)
Mean
(g/L)
Min
(g/L)
Max
(g/L)
SD
(g/L)
1
42
4.04
0.94
15.18
0.047
4.05
0.84
15.64
0.059
2
20
1.47
0.65
2.49
0.066
1.66
0.70
2.75
0.130
3
64
2.54
0.40
6.45
0.122
2.65
0.40
6.70
0.084
The regression line and the coefficient of correlation
Group
N
Regression line
The coefficient
of correlation
Range
Photometers
1
42
1.039 x ICSH–0.150
0.9997
0.94–15.18
5
2
20
1.149 x ICSH–0.024
0.9913
0.65–2.49
1
3
64
1.015 x ICSH–0.078
0.9981
0.40–6.45
1
Содержание Plasma/Low Hb
Страница 1: ...HemoCue Plasma Low Hb Operating manual Manual de funcionamiento Manuel d utilisation ...
Страница 2: ......
Страница 6: ...1 5 4 3 2 6 6 Components Componentes Composants ...
Страница 8: ...1 2 3a 5 3c 4 3b 8 Start up Activación Démarrage ...
Страница 10: ...5 3 4 2 2 1 10 Measuring Sample material Medición Tipo de muestra Mesure Échantillons ...
Страница 12: ...7 8 6 9 12 ...
Страница 14: ...2 1 14 Maintenance Mantenimiento Maintenance ...