27
Specific performance characteristics
The results given below are from a HemoCue Plasma/Low Hb
Photometer standardized against the HiCN-(ICSH)
2
method.
Within-run precision
Within-run precision was determined on 5 photometers and by using
within house prepared aqueous and plasma solutions.
Hemoglobin in 0.9 % NaCl.
Solution
Mean
(g/L)
SD
(g/L)
CV
%
Number
Low
1.59
0.06
3.6
20
Mean
16.30
0.14
0.8
20
High
25.30
0.16
0.6
20
Hemoglobin in heparinized plasma.
Solution
Mean
(g/L)
SD
(g/L)
CV
%
Number
Low
1.95
0.08
4.2
20
Mean
15.30
0.11
0.7
20
High
25.50
0.22
0.9
20
Between-run precision
Between-run precision was determined by using within house prepared
solutions. The sample were analyzed on 20 consecutive working days,
twice a day, in duplicate.
Hemoglobin in 0.9 % NaCl.
Solution
Mean
(g/L)
SD
(g/L)
CV
%
Number
Low
1.43
0.06
4.3
20
Mean
14.80
0.17
1.2
20
High
24.80
0.25
1.0
20
Hemoglobin in heparinized plasma.
Solution
Mean
(g/L)
SD
(g/L)
CV
%
Number
Low
1.78
0.09
5.0
20
Mean
14.80
0.21
1.4
20
High
24.50
0.25
1.0
20
Содержание Plasma/Low Hb
Страница 1: ...HemoCue Plasma Low Hb Operating manual Manual de funcionamiento Manuel d utilisation ...
Страница 2: ......
Страница 6: ...1 5 4 3 2 6 6 Components Componentes Composants ...
Страница 8: ...1 2 3a 5 3c 4 3b 8 Start up Activación Démarrage ...
Страница 10: ...5 3 4 2 2 1 10 Measuring Sample material Medición Tipo de muestra Mesure Échantillons ...
Страница 12: ...7 8 6 9 12 ...
Страница 14: ...2 1 14 Maintenance Mantenimiento Maintenance ...