183
Symptom
Explanation
Action
Docking
Station LED
red
1) Flashing red light - external
communication error.
2) Steady red light - internal
error in the Docking Station.
1) See Symptom “
No transfer of
data
”
.
2a) Disconnect and then reconnect
the Power Adapter.
2b) Contact HemoCue AB or the
distributor.
No transfer of
data via USB
No USB-communication.
a) Check that the Analyzer is
properly docked.
b) Check the cable connections in
both the Docking Station and the
PC.
c) Check that the Docking Station is
not a Primary Docking station.
d) Check that all Docking Stations
that should communicate are
connected with a USB cable to
the PC.
e) Check that the DMS Software is
properly installed and that the
USB driver is installed.
f) Check the electrical connection.
The LED on the docking station
should give a stable or flashing
green light.