157
QC Level 1
Recent Trend
Lot No:
1413-1245
11
15
12
13
14
FIGURE 7-13
The diagram shows the trend for the most
recent QC measurements at the defined
concentration level. No consideration is given
to the defined date interval.
e) Press the Confirm button (15).
11 Lot No. of Liquid Control
12 Fail area
13 Warning area
14 Pass area
15 Confirm button
QC Level 1
All Statistics
Lot No:234567
N= 30
= 52 mg/dL
SD= 0.76 mg/dL
CV= 1.5 %
16
18
19
21
09/10/03 - 09/11/03
17
22
20
FIGURE 7-14
The image shows statistical data for all
stored QC tests with the current Lot No.
f) Press the Confirm button (22).
will be displayed.
16 Lot No. of Liquid Control
17 Date interval of test results
18 Number of test results on which statistics are
based
19 Mean value
20 Standard Deviation
21 Coefficient of Variation
22 Confirm button