111
Patient Test
PID: 123456789
LID: 123456789
357 mg/dL
Result
8
7
2
4
Critical
5
6
1
3
FIGURE 4-12
To add comments to the result, press the
Comment input button (6). See
.
The Verify/Duplicate sampling button (7)
allows the verification of the result by
measuring a new sample from the patient
(see
4.1.2 Verify/Duplicate sampling
Do not press the Confirm button (8) in this
screen if the result is to be verified.
Never re-measure a Cuvette!
a) Press the Confirm button (8) to store the
information.
The Main Menu,
will be
displayed.
The result will remain on the display even if
the Cuvette holder is pulled out, allowing for
examination of the Cuvette before
comments are made.
1 Type of test
2 Patient ID
3 Lab Number ID
4 Result
5 Critical Note
6 Comment input button
7 Verify/Duplicate sampling button
8 Confirm button
Содержание 201 DM
Страница 1: ...HemoCue 201 DM Analyzer HemoCue 201 DM Docking Station Reference Manual ...
Страница 6: ...PART I Introduction ...
Страница 41: ...PART II Settings ...
Страница 108: ...PART III Routine use ...
Страница 181: ...PART IV Other actions ...
Страница 192: ...PART V Miscellaneous ...