Helmer Scientific i.Series
®
Platelet Incubator
Operation Manual
360388/A
10
3.6
Active Alarms
Home screen with active alarm
Notes
• When the door switch is bypassed, the incubator and door open alarm continue to operate as if the door is closed.
• The door switch may be bypassed by opening the door and pulling the switch cylinder.
• The number of agitators, internal and external fans, and heat pumps will vary based on incubator model.
• Fan failure alarm may occur if facility power is lost.
Table 2. i.Series Active Alarms
Alarm
Description
Primary Monitor Probe High Temp
Primary monitor probe reading is above high temperature alarm setpoint
Primary Monitor Probe Low Temp
Primary monitor probe reading is below low temperature alarm setpoint
Primary Monitor Probe Failure
Primary monitor probe is not functioning properly
Control Probe Failure
Control probe is not functioning properly
Agitator 1 Communication Failure
Agitator 1 not installed; communication cable not installed; Agitator On/Off switch turned OFF
Agitator 1 High Speed
Agitator 1 speed (CPM) is above high speed alarm setpoint
Agitator 1 Low Speed
Agitator 1 speed (CPM) is below low speed alarm setpoint
Agitator 1 Maintenance Required
Check and replace trolley support wheels
Condensate Tray Full
Rear condensate tray is full
(recommendation it to empgy soon to avoid averflow)
External Fan 1 Failure
External Fan 1 is not functioning properly
External Fan 2 Failure
External Fan 2 is not functioning properly
Internal Fan 1 Failure
Internal Fan 1 is not functioning properly
Heat Pump 1 High Temp
Heat Pump 1 probe reading is above high temperature alarm setpoint
Heat Pump 1 Connection Error
Temperature is moving away from setpoint
Heat Pump 1 Failure
Heat Pump 1 is not functioning properly
Drive Space is Low
SD card is near capacity
Drive Space is Full
SD card is full, no history being recorded
Power Failure
Power to unit has been disrupted
Door Open (time)
Door is open beyond user-specified duration
Low Battery
Rechargeable battery voltage is low
No Battery
Battery is not connected
MPB Communication Failure
Communication is lost with heat pump controls
Communication Failure Messages 1, 2, 3
1 Communication lost between i.C
3
display board and control board
2 Communication lost between i.C
3
display board and internal system memory
3 Corrupt database