background image

Helmer Scientific i.Series

®

 Platelet Incubator

Operation Manual 

360388/A

17

Preventive Maintenance

Maintenance tasks should be completed according to the following schedule.  Refer to the service manual for more detail on the 

various tasks.

 Note

These are recommended minimum requirements.  Regulations for your organization or physical conditions at your 

organization may require maintenance items to be performed more frequently, or only by designated service personnel.

Task

Frequency

Quarterly

Annually

As Needed

Test the high and low temperature alarms.

ü

Test the power failure alarm (as required by your organization’s protocols).

ü

Test the door open alarm.

ü

Test the no battery alarm.

ü

Test the motion alarm.

ü

Check the temperature calibration for the temperature monitor and change it if 

necessary.

ü

Check the temperature calibration for the temperature controller and change it if 

necessary.

ü

Check the backup battery for the motion alarm system after an extended power 

failure and change it if necessary. (qty 1, 12v 7 AHr SLA battery)

ü

(only PC100-Pro with integrated full system battery backup)

Check the system backup battery functionality to operate the complete system 

(+PF15-Pro agitator) and change if necessary, or change batteries (2) if in service 

for 2 years. (12v 18AHr SLA batteries)

ü

Check the backup battery for the chart recorder after an extended power failure 

and change it if necessary, or change the battery if it has been in service for one 

year. Refer to the Temperature Chart Recorder Operation and Service Manual.

ü

Clean the external fan grill.

ü

Clean the exterior and interior.

ü

Clean the door gaskets.

ü

Replace moving parts if worn or when prompted by the agitation maintenance 

reminder.

ü

 Notes

•  During a power failure, the sealed lead acid (SLA) backup battery provides power to the monitoring system  

  and the power failure alarm. If the backup battery is not functioning, the power failure alarm will not be activated.
•  If the backup battery does not provide power to the monitoring system during the power failure alarm test, replace  

  the battery.
•  Use only a 12V 7AHr SLA battery for the standard monitoring and alarm system battery backup.
•  Use only a 12V 18AHr SLA battery (quantity 2) for the PC100-Pro integrated full system battery backup.

Содержание i.Series PC100-Pro

Страница 1: ...360388 A Platelet Incubator Operation Manual i Series Pro Models Countertop PC100 Pro PC900 Pro PC1200 Pro ...

Страница 2: ...ght 2019 Helmer Inc All other trademarks and registered trademarks are the property of their respective owners Helmer Inc doing business as DBA Helmer Scientific and Helmer Disclaimer This manual is intended as a guide to provide the operator with necessary instructions on the proper use and maintenance of certain Helmer Scientific products Any failure to follow the instructions as described could...

Страница 3: ...on 7 3 1 Initial Start Up 7 3 2 Operation 8 3 3 Users and Passwords 8 3 4 Change Temperature Setpoint 9 3 5 Set Alarm Parameters 9 3 6 Active Alarms 10 3 7 Mute and Disable Active Alarms 11 3 8 Configure a Pro Series Platelet Agitator for Use in a Pro Series Incubator Optional 12 4 Min Max Temperature Monitoring 13 5 Electronic Access Control Optional 14 5 1 Setup 14 5 2 Open Incubator with Access...

Страница 4: ...rect the interference at his own expense Helmer platelet incubators are intended to provide the controlled temperature environment required for the storage of platelet products The devices are intended to be operated by personnel who have procedures in place for meeting FDA AABB EU or any other applicable regulations for the processing and storage of platelet products 1 4 Safety Precautions and Sy...

Страница 5: ...nts when closing the door or when agitation motion is enabled if applicable Avoid sharp edges when working inside the electrical compartment Ensure biological materials are stored at recommended temperatures determined by standards literature or good laboratory practices Proceed with caution when adding and removing product from the platelet incubator Use manufacturer supplied power cord only Usin...

Страница 6: ...s are provided with the unit and should be installed prior to placing the incubator in its location 1 Align keyhole openings in stand off bracket with screws on back of incubator and slide down to engage 2 Tighten screws using a 2 Phillips screwdriver to secure 3 Place platelet incubator on study surface 4 Ensure platelet incubator is level 2 3 Chart Recorder Note For complete information refer to...

Страница 7: ...recorder door 2 4 Install Pro Series Platelet Agitator Optional Connect the data cable and DC power cable supplied with the incubator prior to placing the agitator inside the incubator The agitator communication switch must be set to position 1 for use in the incubator For additional information refer to the Platelet Agitator Operation Manual Note Only Helmer Pro series platelet agitator may be us...

Страница 8: ... specification label 2 Turn the AC power switch ON 3 Turn the backup battery switch ON 4 The Start screen is displayed Start screen The Language screen is displayed following the Start screen Use the Language drop down menu to select the i C3 display language Language screens Note English is the default language If an alarm sounds temporarily mute the alarm by touching the Mute icon Home screen Mu...

Страница 9: ... provides a way for the administrator level user to limit access to certain screens The administrator level password can be changed by selecting Change Password The Access Control Setup screen can be opened by choosing the Access Setup button from the Users and Passwords screen as well as from the Access Log screen Enter settings password Select Users and Passwords Users and Passwords screen Chang...

Страница 10: ...box to change value Settings screen Notes Default Settings password is 1234 Default setpoint is 22 0 C 3 5 Set Alarm Parameters Enter Settings password Select Alarm Settings Touch minus or plus on spin box to set each alarm parameter Settings screen Alarm Settings screen Alarm settings control the circumstances and timing of alarm condition indicators displayed on the i C3 Home screen ...

Страница 11: ...high speed alarm setpoint Agitator 1 Low Speed Agitator 1 speed CPM is below low speed alarm setpoint Agitator 1 Maintenance Required Check and replace trolley support wheels Condensate Tray Full Rear condensate tray is full recommendation it to empgy soon to avoid averflow External Fan 1 Failure External Fan 1 is not functioning properly External Fan 2 Failure External Fan 2 is not functioning pr...

Страница 12: ...iption Icon Description Icon Description Icon Description Home Alarm Test Download Save Event Log Information Logs CSV Download Cancel Mute AgiTrak PDF Download Back Arrow Reset Contact Information Contact Helmer Access Control Scroll Zoom Information Display Brightness Access Log Temperature Graph Forward Back i C Applications Icon Transfer Alarm Conditions Zoom Out Settings Upload Cancel Test Ba...

Страница 13: ...rough the AgiTrak system when an agitator is configured inside an incubator This setting overrides the setting on the agitator speed control Helmer recommends the agitator alarm remain enabled when configured inside an incubator With the alarm enabled the agitator alarm will time out and sound if power to the incubator is turned off for a duration greater than the motion delay In the event of a co...

Страница 14: ...ure reading since the last system reset power on event or manually initiated reset Touch the Reset icon to the right of the display to manually reset Notes The Min Max temperature display can be turned on or off through Display Settings Once the time reaches the maximum display of 999 hours and 60 minutes the message will display 999 60 but minimum and maximum temperatures will continue to be trac...

Страница 15: ...er failure During a power failure use the mechanical door key to provide secure storage for incubator contents Refer to i C3 User Guide for complete information regarding Access Control 5 1 Setup Configure and manage user specific accounts to allow controlled access to the incubator Access Control with Password overlay Access Control Setup screen Enter the supervisor PIN to set up Access Control F...

Страница 16: ...Hz 60 Hz Voltage Tolerance 10 Circuit Breakers 7 0A quantity 2 100V units quantity 1 Agitator Power Draw if installed 16 W at 24 V DC 16 W at 24 V DC 16 W at 24 V DC 16 W at 24 V DC Incubator Power Consumption with agitator installed 65 Watts 352 Watts 75 Watts 415 Watts 65 Watts 352 Watts 65 Watts 352 Watts Power Source Varies refer to product specification label Remote Alarm Capacity 1A at 33V A...

Страница 17: ...When disposing of this product in countries affected by this directive Do not dispose of this product as unsorted municipal waste Collect this product separately Use the collection and return systems available locally For more information on the return recovery or recycling of this product contact your local distributor 7 3 Electromagnetic Compliance Helmer Scientific Incubators meet the applicabl...

Страница 18: ... SLA battery ü only PC100 Pro with integrated full system battery backup Check the system backup battery functionality to operate the complete system PF15 Pro agitator and change if necessary or change batteries 2 if in service for 2 years 12v 18AHr SLA batteries ü Check the backup battery for the chart recorder after an extended power failure and change it if necessary or change the battery if it...

Страница 19: ...nterface J Remote alarm interface C Control panel K RJ45 port D Manual Electronic Access Control override key switch L Circuit breaker s E USB port M AC input power cord receptacle F Back up battery switch N External electronic access control terminal activation optional G Power switch O Condensate pan drain port H Control and chart recorder probe A E C B I H G F PC900 Pro PC1200 Pro M L J N K PC1...

Страница 20: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA Copyright 2019 Helmer Inc 360388 A ...

Отзывы: