360142-D/D
21
i.Series® Models
1
Lay the door assembly face-down on a solid work surface.
2
Remove remaining screws from both sides of the door assembly.
3
Lift the inner door frame out of the outer door frame and set aside. A J-hook tool may be used along
the bottom edge of the door assembly to lift the inner frame.
4
Remove the plug from the handle-side of the door. Set the plug aside.
Pull plug out
5
Pull the grommet out of the hole in the door. Slide the braided sleeve out of the slot.
Pull grommet
out
Slide sleeve
out of slot
6
Re-route the power cable and communication cables.
a
Route cables inside the door; along the opposite side of the door frame; and out through the slot
in the corner opposite from where the cables had previously exited the door.
b
Cables should follow the edge of the door frame, as closely as possible.
c
Tape the cables to the door frame.
d
Excess slack in the cables should be outside of the door.