360142-D/D
40
i.Series® Models
9.2
Chamber Temperature Problems
Problem
Possible Cause
Action
The chamber
temperature displayed
is higher or lower
than the actual
temperature.
Monitor probe is not
calibrated.
►
Check the chamber temperature calibration.
Change the calibration if necessary.
Connections for the
monitor probe are
loose.
►
Test the monitor probe connections. Secure the
connections if necessary.
Monitor probe wiring
is an open circuit.
►
Check the continuity of the probe wiring. Replace
the probe if necessary.
Probe bottle is empty,
or the amount of
solution is too low.
►
Check the level of product simulation solution in
the bottle. Clean and refill the bottle if necessary.
Solution in the probe
bottle is frozen.
►
Refill the bottle with new solution.
Digital electronics are
locked because of an
interruption in power.
►
Reset the monitoring system.
Monitor is not
calibrated.
►
Confirm the monitor probe is reading correctly.
Calibrate the monitor probe if necessary.
The chamber
temperature meets
an alarm condition,
but the appropriate
temperature alarm is
not active.
Temperature alarm
setpoint was changed.
►
Check the current setpoints for the temperature
alarms. Change the setpoints if necessary.
The compressor runs
continuously.
Freezer setpoint is set
too low.
►
Confirm the setpoint is set within the operating
range and change it if necessary.
Temperature monitor/
control board is faulty.
►
Confirm the temperature controller or monitor/
control board is operating correctly. Replace it if
necessary.
Compressor starting
relay is faulty.
►
Confirm the relay is operating correctly. Replace
the relay if necessary.
Содержание Horizon HLF105
Страница 56: ...360142 D D 55 i Series Models 11 Schematics 11 1 iPF and iLF Models 105 Configuration ...
Страница 57: ...360142 D D 56 i Series Models Monitor Control Control ...
Страница 98: ......