360142-D/D
65
Horizon Series™ Models
Label Description
A
Outer door frame
B
Data cable (gray)
C
Cable exit (corner of door)
D
Additional cable slack
1
Lay the door assembly face-down on a solid work surface.
2
Remove remaining screws from both sides of the door assembly.
3
Lift the inner door frame out of the outer door frame and set aside. A J-hook tool may be used along
the bottom edge of the door assembly to lift the inner frame.
4
Remove the plug from the handle-side of the door. Set the plug aside.
Pull plug out
5
Pull the grommet out of the hole in the door. Slide the data cable out of the slot.
Pull grommet
out
Slide cable
out of slot
6
Re-route the data cable.
a
Route the data cable inside the door and out through the slot in the corner opposite from where
the cable had previously exited the door.
b
Cable should follow the bottom edge of the door frame, as closely as possible.
c
Tape the cable to the door frame.
d
Excess slack in the cable should be outside of the door.