Maintenance
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
65
Be
for
e
You Beg
in
In
stallation
A
na
log O
per
atio
n
R
S–48
5 Op
er
ation
M
ain
te
n
a
n
ce
Table 5-2
Troubleshooting RS–485 error messages
Message from
module
Possible causes
Solution
DEF SNSR
• Convectron gauge is defective.
• Diaphragm pressure sensor is defective.
• Pressure is > 1000 Torr.
• Process gas is a not the gas that the user
anticipated using in the system.
• If Convectron gauge or diaphragm pressure
sensor is defective, return module to factory (see
page 70).
• If pressure is higher than 1000 Torr, decrease
pressure.
• If process gas is not the gas that the user
anticipated, adjust relays indicating vacuum
pressure for gas that is being used (see page 56).
HI ATM V
• Atmospheric pressure is > pressure represented
in TS command.
• TO command was sent while atmospheric
pressure < 400 Torr (5.332 mbar, 533.2 Pa)
• Convectron gauge is contaminated.
• Module is too near a gas inlet.
• If atmospheric pressure is too high, decrease
pressure, then recalibrate (see page 56).
• If atmospheric pressure is < 400 Torr
(5.332 mbar, 533.2 Pa), increase pressure, then
recalibrate differential pressure sensor zero (see
page 57).
• If Convectron gauge is contaminated, return it to
factory (see pages 68 and 70).
• If module is too close to a gas inlet, relocate
module farther away from inlet (see page 7).
LO ATM V
• Atmospheric pressure is < pressure represented
in TS command.
• TO command was sent while atmospheric
pressure was > high limit of measurement range.
• Module is too near pump.
• If atmospheric pressure is too low, increase
atmospheric pressure, then recalibrate
Convectron gauge (see page 56).
• If atmospheric pressure was too low, increase
pressure, then recalibrate differential pressure
sensor zero (see page 57).
• If module is too near pump, relocate module
farther away from pump (see page 7).
GAIN LIM
TS command was sent while atmospheric pressure
was > high limit of measurement range.
Decrease atmospheric pressure, then recalibrate
Convectron gauge (see page 56).
HI VAC V
Vacuum pressure is > pressure represented in TZ
command.
Decrease vacuum chamber pressure, then
recalibrate Convectron gauge (see page 56).
LO VAC V
Vacuum pressure is < pressure represented in TZ
command.
Increase vacuum chamber pressure, then
recalibrate Convectron gauge (see page 56).
OFST LIM
• TZ command was sent while vacuum chamber
pressure was < low limit of measurement range.
• TO command was sent while differential
pressure was outside measurement range.
• If vacuum chamber pressure < low limit of
measurement range, increase vacuum chamber
pressure, then recalibrate Convectron gauge (see
page 56).
• If differential pressure was outside measurement
range, increase pressure, then recalibrate
differential pressure sensor zero (see page 57).
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......