Analog Module Operation
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
27
Be
for
e
You Beg
in
In
stallation
A
na
log O
per
atio
n
R
S–48
5 Op
er
ation
M
ain
te
n
a
n
ce
Figure 3-8
Relay 2 behavior
Setpoint 1: Convectron
gauge
If the module will measure the pressure of any process gas other than N
2
or
air, you must adjust setpoint 1 for the process gas that will be used.
The true pressure of a gas other than N
2
or air may be substantially different
from the pressure that the output indicates. For example, outputs might
indicate a pressure of 10 Torr for argon, although the true pressure of the
argon is 250 Torr. Such a substantial difference between indicated pressure
and true pressure can cause overpressurization resulting in an explosion.
Test points on the module front panel, Table 3-5 on page 22, and
Figure 3-10 and Figure 3-11 on pages 29–30 enable you to determine the
amount of voltage that corresponds to a specific pressure.
1.
Make sure the module is properly installed, the axis is horizontal (see
page 8), and the power supply is ON.
2.
Figure 3-10 and Figure 3-11 are graphs that show test point voltage
versus pressure for N
2
or air. Table 3-5 on page 22 lists test point
voltage versus pressure for 12 commonly used process gases, including
N
2
and air.
Activate
Deactivate
Deactivated
Activated
Relay
activated
8 Torr
hysteresis
Time
Differential pressure
WARNING
Failure to use accurate pressure conversion data for N
2
or air to other gases can cause an explosion due to
overpressurization.
If the module will measure any gas other than N
2
or air, before
connecting relays to system control devices, adjust setpoints for
the process gas that will be used.
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......