Specifications
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
77
Specifications for RS-485 version
Digital format
American Standard Code for Information Interchange (ASCII)
Protocol
RS–485, 2-wire
Parity
• 8 data bits, no parity, 1 stop bit (default)
• 7 data bits, odd parity, 1 stop bit
• 7 data bits, even parity, 1 stop bit
Baud rate
1200, 2400, 4800, 9600, 19200 (default), or 38400 baud
Setpoint relays
Four single-pole, single-throw (SPST) relays. Use RS–485 commands to
configure relays. Each relay can be assigned to the Convectron gauge or the
differential pressure sensor.
Relays 1 and 3
• In default mode, relays 1 and 3 are assigned to the Convectron gauge,
which indicates vacuum chamber pressure
• Range: 10
–3
to 1000 Torr
Relays 2 and 4
• In default mode, relays 2 and 4 are assigned to the differential pressure
sensor, which indicates pressure differential between atmospheric and
vacuum chamber pressures
• Range: –750 to +125 Torr
RS–485 male I/O connector on module end panel
No connection
5
Power ground
4
11–26 V power input
3
RS–485 B line
2
RS–485 A line
1
15
Relay 1 common
14
Relay 4 common
13
No connection
12
Relay 2 common
11
Relay 2 normally open
10
Relay 1 normally open
9
Relay 4 normally open
8
RS–485 ground
7
Relay 3 common
6
Relay 3 normally open
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......