Chapter 3
32
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
2.
Use a high-precision, high-input impedance (Z
in
> 1 M
Ω
) DVM to
measure the setpoint voltage across the SETPOINT2 and COMMON
test points on the module front panel. Figure 3-13 illustrates
SETPOINT2 and COMMON test points.
3.
Use a flat-head instrument screwdriver to adjust the SETPOINT2
potentiometer for the desired setpoint voltage at which relay 2 will
activate and deactivate. (See Figure 3-13.)
A built-in 8 Torr hysteresis prevents oscillation around the setpoint.
Figure 3-13 Test points and potentiometer for setpoint relay 2
COMMON test point
SETPOINT2 test point
SP2 ADJ potentiometer