Installation
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
7
Be
for
e
You Beg
in
In
stallation
A
na
log O
per
atio
n
R
S–48
5 Op
er
ation
M
ain
te
n
a
n
ce
Step 1
Install pressure relief devices
Before you install the module, install appropriate pressure relief devices in
the vacuum system.
Helix Technology does not supply pressure relief valves or rupture disks.
Suppliers of pressure relief valves and rupture disks are listed in the
Thomas
Register
under “Valves, Relief” and “Discs, Rupture.”
Step 2
Locate and orient module
To locate and orient the module, refer to Figure 2-3 on page 8 and follow
the instructions below.
Locate the module
• For greatest accuracy and repeatability, locate the module in a stable,
room-temperature environment. Ambient temperature should never
exceed 40 °C (104 °F) operating, non-condensing, or 85 °C (185 °F)
non-operating.
• Locate the module away from internal and external heat sources and in
an area where ambient temperature remains reasonably constant.
• Do not locate the module where it requires long lengths of tubing or has
constricted tubing. Length of tubing depends on the application. Longer
lengths will affect vacuum pressure limit and response time.
• Do not locate the module near the pump, where gauge pressure might be
lower than normal vacuum chamber pressure.
• Do not locate the module near a gas inlet or other source of
contamination, where inflow of gas or particulates causes atmospheric
pressure to be higher than system atmosphere.
• Do not locate the module where it will be exposed to corrosive gases
such as mercury vapor or fluorine.
• Do not locate the module where it will vibrate. Vibration causes
convection cooling, resulting in inaccurate high pressure readings.
CAUTION
Operating the module above 1000 Torr (1333 mbar, 133
kPa) true pressure could cause pressure measurement
error or product failure.
To avoid measurement error or product failure due to
overpressurization, install pressure relief valves or rupture
disks in the system if pressure exceeds 1000 Torr (1333 mbar,
133 kPa).
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......