Chapter 2
8
Convectron
®
ATM Vacuum Gauge Module Instruction Manual
Orient the module
For proper operation of the module above 1 Torr, orient the module so the
axis is horizontal (see Figure 2-3). Although the Convectron gauge will read
correctly below 1 Torr with the module mounted in any position, inaccurate
readings will result at pressures above 1 Torr if the module axis is not
horizontal.
Figure 2-3
Module orientation
Mount module axis horizontally to ensure
accurate measurement above 1 Torr
Vacuum chamber
Vacuum chamber
Va
cuum chambe
r
V
acuum chamber
V
ac
uu
m
ch
am
be
r
V
ac
uu
m
Vacuum
Recommended
Not recommended