Convectron
®
ATM Vacuum Gauge Module Instruction Manual
5
Be
for
e
You Beg
in
In
stallation
A
na
log O
per
atio
n
R
S–48
5 Op
er
ation
M
ain
te
n
a
n
ce
Chapter 2
Installation
2.1
Module components
The Convectron ATM module contains a Convectron convection-enhanced
Pirani heat-loss gauge and a Piezo resistive diaphragm sensor.
The module is shipped with an instrument screwdriver and a 15-pin female,
high-density subminiature D connector that mates to the male connector on
the module.
Figure 2-1 illustrates the analog version of the module. In the analog
version, setpoint relay 1 is assigned to the Convectron gauge, and setpoint
relay 2 is assigned to the differential pressure sensor.
Figure 2-1
Analog version of module
WARNING
Using the module to measure the pressure of flammable
or explosive gases can cause a fire or explosion resulting
in severe property damage, personal injury, or death.
Do not use the module to measure the pressure of flammable or
explosive gases.
Vacuum chamber
fitting
Test points
Setpoint activation
indicators
15-pin male
subminiature D
connector
Calibration and
setpoint potentiometers