Transmission/Reflection Mode (Setting of transmittance)
5.2.6.3.
Setting of transmittance indicates to what extent the ambient temperature is
transmitted through the measured target and becomes a part of the received re-
flection.
Emissivity (
𝜀
) is calculated by reflection (
𝜌
) and transmission (
𝜏
) by formula
ε = 1 − ρ − τ
.
Call up menu
Select menu item
Confirm
Select menu item
Use selection
Display and setting of reflection
and transmission values (if
activated)
Code setting see 5.2.4
Select menu item
Change value
Select digit
Change value
Save value
Further options (see
Option Intern/Extern/Manual)
Press several times or keep
pressing to return to measure-
ment