79
Percent Error Operation messages
• % Err: in & out match: refer to section
• % Err: bad input range: refer to section
• % Err: bad output range:
• % Err: ext input of: press V/mA key to turn external channel on
• % Err: CH1 empty: no module is installed in Channel 1 bay
• % Err: CH2 empty: no module is installed in Channel 2 bay
Dual Sensor Operation messages
•
DS err: CH1 unplugged: no module is installed in Channel 1
bay
•
DS err: CH2 unplugged: no module is installed in Channel 2
bay
•
DS err: units mismatch: engineering units of installed modules
are different. Use units MENU to make both modules display
same engineering units.
Switch Test Operation messages
• ST err: CH1 unplugged: no module is installed in Channel 1
bay
•
ST err: CH2 unplugged: no module is installed in Channel 2
bay
•
Sw. Test forced V in: Voltage input was not enabled prior to
initiating switch test. The HHC has automatically enabled
voltage input monitoring
•
Sw. Test forced mA in: mA input was not enabled prior to
initiating switch test. The HHC has automatically enabledmA
input monitoring
• No 24v loop (IS unit):
Appendix C
ISO 1745 Commands
Format for command
ASCII
Included
non-
(hex)
in BCC =
printable (hex)
xor’d
SOH
01
DLE
10
STX
02
Y
<cmd>
Y
HHC command
DLE
10
ETX
03
Y
BCC
BCC = byte[2] ^ byte[3] ^ byte[5]
ASCII
code
hex
SOH
01
DLE
10
STX
02
ETX
03
NAK
15
ACK
06
PTE-2 Response is the same format proceeded by an ACK, a
NACK is sent alone if the command is not understood.
SECTION 16 b.
TROUBLE SHOOTING CONT.
SECTION 16 c.
ISO COMMANDS