CALIBRATION
Calibration
This scale has been factory calibrated, and does not require calibration prior to use.
For
maintenance, the scale can be calibrated. Calibration of your 2400KL scale is performed using
kilograms (kg) or pounds (lb), according to the units of measure used upon entering into the
calibration path. The following calibration assumes the scale is set to pounds, and 400 lb (the
default weight) is employed as the calibration weight.
Scale Calibration
Operator Action
Display
1. Make sure the display is turned off, then press and hold the
HOLD/RELEASE
button while pressing the
ON/OFF
button.
“U 01” followed by a 5 digit
‘idle scale count’ number.
2. Press the
ENTER
button.
“CAL”.
3. Make sure there is nothing on the scale, then press the
ENTER
button again.
“Clear” followed by
“0400.0LB”.
4. Use the
►
and
▲
buttons to set the calibration weight to a
value other than the default weight. The calibration weight should
be as high as possible, with 200 lb as a minimum calibration
weight.
Shows “400.0” or entered
calibration weight.
5. Press the
ENTER
button.
Alternating “Load” followed
by “400.0LB”.
6. Load the scale with the required weight.
Display increments showing
raw weight data.
7. Without touching or shaking the scale, allow the display to
stabilize, then press the
ENTER
button.
“SUCC” or “Err1”.
8. If the calibration is successful, the display will show SUCC and
turn off. If the calibration is not successful, the display will show
Err1 and turn off. Contact customer service for an Err1 display
(+1 800-638-3722).
-------
.
Содержание 2400KL
Страница 1: ...1 Model 2400KL Digital Portable Wheelchair Scale User Instructions P N UM2400KL Rev 1 032212 ...
Страница 9: ...9 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 1 Figure 2 ...
Страница 10: ...10 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 3 Figure 4 Load cell port ...
Страница 11: ...11 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 5 Figure 6 12mm from the surface Folding hinges ...
Страница 12: ...12 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 7 ...
Страница 16: ...16 SETUP BATTERIES TRANSPORT CONTINUED Figure 10 Figure 11 ...