background image

10 

2400KL ASSEMBLY INSTRUCTIONS (CONTINUED) 

 

 

 

Figure 3 

 

 

 

Figure 4 

Load cell port 

Содержание 2400KL

Страница 1: ...1 Model 2400KL Digital Portable Wheelchair Scale User Instructions P N UM2400KL Rev 1 032212 ...

Страница 2: ...6 Assembly Instructions 2400KL Scale 7 Display Assembly Wall Mounting Instructions 13 Setup Batteries Transport 14 Quick Start Instructions 17 Maintenance 21 Calibration 22 Troubleshooting 23 Warranty 24 Note This scale has been factory calibrated and does not require calibration prior to use Please register your scale for warranty coverage at http www homscales com customer registration aspx For ...

Страница 3: ...n damage the equipment and will void the warranty Do not transport the scale with any weight on the platform Do not exceed recommended weight limit of 800 lb 360 kg for this scale For accurate weighing this scale must be placed on a flat stable surface Also verify before each use the proper operation according to the procedure described in this manual Do not use in the presence of flammable or exp...

Страница 4: ...ended throughout the entire weighing event Exercise caution when rolling patient onto and off of the scale To prevent injury exercise care when adjusting wheelchair ramps To prevent injury and damage to your scale do not drop or mishandle scale during handling or transport ...

Страница 5: ...ams kg For privacy and convenience in use the scale features a remote display head which may be held by the operator placed on a table or other flat surface or mounted on a wall Each scale can be powered using an AC adapter included or by 6 AA batteries not included 2400KL Specifications Capacity and Resolution 800 lb x 0 2 lb 360 kg x 0 1 kg Power Requirements 120 240VAC 50 60Hz 9V DC adapter or ...

Страница 6: ...chnical Support available during normal business hours at 1 800 815 6615 or 1 708 377 0600 2400KL Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or local regulations which apply to you for disposal of electronic waste or batteries Do not dispose of this device in the domestic waste stream Pelstar LLC has been offi...

Страница 7: ...y remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Parts List 1 Wheelchair ramp scale 1 Small brown box The small brown box contains 1 Display assembly 1 Load cell cable 1 User information 1 Long whi...

Страница 8: ...se the scale by adjusting all eight feet see Figure 6 8 Connect the load cell cable to the wheelchair ramp by inserting the load cell cable connector into the load cell port located on the inner side of the right ramp see Figure 4 9 If the AC adapter will be used connect the AC adapter cord plug into the AC power jack located on the back of the display assembly The AC power jack is labeled POWER s...

Страница 9: ...9 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 1 Figure 2 ...

Страница 10: ...10 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 3 Figure 4 Load cell port ...

Страница 11: ...11 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 5 Figure 6 12mm from the surface Folding hinges ...

Страница 12: ...12 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 7 ...

Страница 13: ...om each side of the display outline 4 On the level horizontal line mark two drill holes 2 57mm away from the center line 5 Use the appropriate plastic or metal anchors for the type of wall construction and carefully drill the appropriate sized holes 6 Insert the anchors into the wall and screw in the two mounting screws into the anchors Adjust the screw head depth such that the bracket can be moun...

Страница 14: ...battery power if batteries are installed Note If the set up procedure failed refer to the troubleshooting instructions If the problem is not corrected refer to qualified service personnel Replacing Inserting Batteries 1 Unplug the AC adapter from the power source 2 Remove the battery cover on the side of the display assembly by unscrewing the 2 battery cover screws see Figure 11 3 Replace or insta...

Страница 15: ... and remove the adapter from the outlet Carefully store the AC adapter and cord to prevent damage during transport 3 With the scale on the floor fold both ramps by pulling up on the center where the handle is located 4 Position the display assembly onto the mounting pins located on the center of the folded ramp The display assembly is mounted onto the ramp by sliding the display assembly brackets ...

Страница 16: ...16 SETUP BATTERIES TRANSPORT CONTINUED Figure 10 Figure 11 ...

Страница 17: ...t data to calculate patient s Body Mass Index BMI RECALL Recalls the previous weight TARE PRETARE TARE Tares out weight of an object on the platform PRETARE Prompts entry of a tare value that will be deducted from the weight on the platform Also releases tare weight returns display to zero UNIT Toggles between pounds and kilograms Also used to increment height value and pretare value REWEIGH Allow...

Страница 18: ...deduct the tare weight of the wheelchair from the gross weight of the patient and wheelchair together 7 The weight of the patient will appear on the scale 8 With no weight on the scale the tare value will remain on the display until changed cleared or until the scale is turned off 9 To remove the Tare press the TARE button again The word TARE will disappear from the bottom of the display and the t...

Страница 19: ... determined Depending on the movement of the patient on the scale it may take several seconds for the scale to lock onto the weight Caution To prevent patient injury the patient must be attended throughout the entire weighing event Exercise caution when rolling patient onto and off of the scale 5 Once a weight has been displayed while the patient is still on the scale Health o meter recommends per...

Страница 20: ...ll lock on the patient s weight when the stable weight of the patient is determined Depending on the movement of the patient on the scale it may take several seconds for the scale to lock onto the weight 5 Once a weight has been displayed while the patient is still on the scale Health o meter recommends performing a second weighing by pressing the REWEIGH button 6 After obtaining the patient s wei...

Страница 21: ...ce of the total scale for any obvious damage wear and tear 2 Inspect the AC adapter for cord cracking or fraying or for broken bent prongs Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective operation Caution Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Clean all external surfaces with a clean damp cloth or tissue Mil...

Страница 22: ...t number 2 Press the ENTER button CAL 3 Make sure there is nothing on the scale then press the ENTER button again Clear followed by 0400 0LB 4 Use the and buttons to set the calibration weight to a value other than the default weight The calibration weight should be as high as possible with 200 lb as a minimum calibration weight Shows 400 0 or entered calibration weight 5 Press the ENTER button Al...

Страница 23: ...is not placed on a level floor 3 Place the scale on a level floor and begin weighing process again 4 Scale is out of calibration 4 Check weight with known weight value 5 Improper Tare 5 Place the item to be tared on the scale Press REWEIGH Once the weight of the item is displayed press TARE Place the patient and the tared item back on the scale Press REWEIGH again Weighing is performed but the wei...

Страница 24: ...ransit or unauthorized repair or alternations Further the warranty does not cover natural disasters such as fire flood hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty Service make sure you keep your sales receipt or document...