Temperature: Temperature set for a temperature point.
Time: Duration for a temperature point
Temperature variation: Increase temperature variation for each cycle, which will be reflected to
the point in each cycle.
Gradient Width: Modular gradient can be used, and the whole module temperature will be
distributed in gradient.
Time Variation: Increase time variation for each cycle, which will be reflected to the point in each
cycle.
Heating Rate: Change heating rate, and the default is Fastest
Cooling Rate: Change cooling rate, and the default is Fastest
Collect Fluorescence: Mark it to collect fluorescence at current temperature point. Only one
temperature point can be specified for each test. Set it by referring to related agent instructions.
Add at the end: Check it to add the temperature point to the end of the program.