13
S9M
A3019-5.0 en/de/fr
HBM
Rigid
force transfer
Rigid
force transfer
correct
incorrect
Fig. 5.1
Transducer orientation during mounting
NOTE
Please note the maximum permissible loading capacity of the mounting parts
used and of the tension/compression bars, screws and knuckle eyes.
5.3
Mounting the S9M
5.3.1 Mounting with tension/compression bars
In this mounting variant, the transducer is mounted with tension/compression
bars on a construction element and can then be measured in the tensile and
compressive directions. Alternating loads are also correctly recorded if the
transducer is mounted without axial play. For dynamic alternating loads, the
upper and lower threaded connectors must be pre-stressed to above the
maximum operating load and then locked in place.
1. Installation and locking with initial stress (recommended mounting variant):
•
Screw in the threaded connector
•
Pre-stress transducer to 110 % operating load in the tensile direction
•
Hand-tighten locknut
•
Relieve the transducer