Graphical User Interface
FS42
A4250-5.0
HBM: public
61
has its own operation range. It is not possible to con
figure sensors with overlapping operation ranges.
Nevertheless, since these ranges are dynamic, they
will follow the measured wavelength (λ). If a range
from a neighbor sensor overlaps the sensor's range,
measurement errors will occur.
Important
Dynamic ranges should be carefully selected in order to
avoid overlap during measurement. If this happens, read
ing errors will occur.
5
The Formula box allows the definition of the calcula
tions to be performed on the measured value of wave
length shift (x=Δλ=λ-λ
0
) (x=Measured WL-CWL). Here
it is possible to set a formula manually. To do so, click
on the formula box and a formula window will appear
(
).
Fig. 6.16
Formulas can be inserted using Excel syntax on the for
mula area (number
1
in
).