Measuring Examples
42
A4250-5.0
HBM: public
FS42
Tip
A connector is more prone to degradation in applications
where you are constantly connecting and disconnecting
sensors. To avoid early degradation and protect the inter
rogator connector use a small patch cable and an
adapter.
5.4.3
Reflective Fiber Ending
Even when all connections have been performed cor
rectly, it may occur that the interrogator does not find any
sensor on the array. One of the reasons may be the
excess of reflected light saturating the detectors.
When a fiber is perfectly cut (
amount of light is reflected right back to the fiber core in
the right direction, heading to the interrogator detectors.
When this occurs, an index matching gel or an angled
termination must be used. If these are unavailable, the tip
of the fiber can be smashed with a metallic tool. This will
destroy the perfect geometry (
on the right) and
the light that is reflected will take random directions end
ing up outside of the core.
Содержание FS42
Страница 1: ...User Manual English FS42 Portable BraggMETER...
Страница 11: ...Hard Disk Image Recovery FS42 A4250 5 0 HBM public 11 7 Press Next Fig 2 4 Fig 2 4 8 Press Yes Fig 2 5 Fig 2 5...
Страница 24: ...Operation 24 A4250 5 0 HBM public FS42 Fig 4 7...
Страница 29: ...Operation FS42 A4250 5 0 HBM public 29 Fig 4 11...
Страница 30: ...Operation 30 A4250 5 0 HBM public FS42 Fig 4 12...
Страница 31: ...Operation FS42 A4250 5 0 HBM public 31 Fig 4 13...
Страница 44: ...Measuring Examples 44 A4250 5 0 HBM public FS42 Fig 5 11...
Страница 74: ...Graphical User Interface 74 A4250 5 0 HBM public FS42...
Страница 75: ...Graphical User Interface FS42 A4250 5 0 HBM public 75...