Measuring Examples
FS42
A4250-5.0
HBM: public
39
5.4
Common Measuring Difficulties
There may be the possibility of having measuring mal
functions due to several problems in the sensing network
or optical connections. A list of some of the usual prob
lems is described below with the explanation for its
occurrence and correction.
5.4.1
Dirty Connector
When an interrogator is repeatedly being plugged in and
out with optical connectors, it is very important that the
connectors are cleaned prior to any connection. Other
wise, dust and moister can be deposited in the interroga
tor adaptor and this will compromise measurements. In
a picture of a magnified connector is presented.
The dark gray circle corresponds to the fiber cladding
and the small light gray circle is the core of the fiber. One
picture for a clean connector and one picture for a dirty
connector are presented.
Clean connector
Dirty connector
Fig. 5.6
The most common effect of dirt on the connections is a
large amount of broad band light being reflected at the
connection, in both directions, meaning that the dynamic
range for measurements becomes smaller.
Содержание FS42
Страница 1: ...User Manual English FS42 Portable BraggMETER...
Страница 11: ...Hard Disk Image Recovery FS42 A4250 5 0 HBM public 11 7 Press Next Fig 2 4 Fig 2 4 8 Press Yes Fig 2 5 Fig 2 5...
Страница 24: ...Operation 24 A4250 5 0 HBM public FS42 Fig 4 7...
Страница 29: ...Operation FS42 A4250 5 0 HBM public 29 Fig 4 11...
Страница 30: ...Operation 30 A4250 5 0 HBM public FS42 Fig 4 12...
Страница 31: ...Operation FS42 A4250 5 0 HBM public 31 Fig 4 13...
Страница 44: ...Measuring Examples 44 A4250 5 0 HBM public FS42 Fig 5 11...
Страница 74: ...Graphical User Interface 74 A4250 5 0 HBM public FS42...
Страница 75: ...Graphical User Interface FS42 A4250 5 0 HBM public 75...