HB-CL2
Cleaning Unit
Operation
50
O8305-EN 2022-01
8.1.2.1 Cleaning phase
Check seal / add cleaning agent
Fig. 20: Warning medium
Proceed as follows if the warning
Add agent
is displayed:
1.
Acknowledge horn with the
key.
2.
Proceed to check the seal in the external circuit.
If a leak appears, cancel the cleaning cycle and switch off the
unit with the
key.
3.
Fill indicated cleaning agent quantity in through the tank
opening in the tank.
WARNING!
Risk of injury due to chemicals!
Chemicals can cause chemical burns, depending
on the type and dilution, cause irritation of the
membrane and respiratory system and be toxic if
swallowed.
Therefore:
– Only allow work on with chemicals s to be
carried out by qualified personnel.
– The general safety instructions on how to
handle chemicals according to the safetydata
sheets must be observed at all times.
– Do not mix chemicals
4.
Acknowledge after adding of cleaning agent the query with the
key.
Cleaning operation
Fig. 21: Set cleaning duration
After having acknowledged the warning
Add agent
the cleaning
operation starts automatically. In this phase the agent is pumped
through the components to be cleaned.
Set if necessary the desired maximum cleaning duration:
1.
Display the menu page
Settings \ Cleaning
.
2.
Set parameter
Cleaning time
to the desired value.
NOTICE!
The unit calculated permanently the cleaning rate
through the pressure change. If the cleaning rate is
steady through a longer period of time, the unit
stops automatically the cleaning operation,
independently of the set cleaning duration.