HATCH X-RAY Operating Manual
49
CHAPTER
7
TROUBLESHOOTING
Hatch X-Ray has been designed with safety features to protect the patient and
operator in case of failure of an electrical component. The system automatically
checks for errors and will report a malfunction by means of an error code on the
Keypad Console display. These error codes are listed in the”Table 11” below. The
table lists the code and identifies the corresponding equipment operational fault.
The corrective action for the error code is provided by “Table 12”, which also lists
observed problems and their potential source.
Error Code
Error
CN001
Communication error
CN002
Console and tube-head are incompatible
CN003
X-Ray preparation time-out
CN004
Anode arc fault
CN005
Cathode arc fault
CN006
Over KV fault
CN007
Over mA fault
CN008
KV regulation fault
CN009
Filament open fault
CN0010
Filament limit fault
CN0011
CAN fault
KB001
Key jam error
Table 11 : Error codes
Содержание X-RAY
Страница 1: ...OPERATING MANUAL ...
Страница 3: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 37: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 45: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 57: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 59: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 65: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...