HATCH X-RAY Operating Manual
42
Attention/Warning
Messages
Screen
Condition
ATTENTION
No modification made to
settings.
Config Accessories
SET key pressed on
“Save and return” option
without modifying any
accessories.
ATTENTION
Press UP/DOWN to
scroll or SET to accept.
Config Accessories
Any one of the following
keys pressed.
1) Mode
2) EXPOSURE
3) Exposure Access or
4) SEL
INFO
Start-up mode already
saved.
Home
SET key pressed when
Current combination
of exposure mode
and tooth anatomy is
the default start-up
combination.
ATTENTION
Not available in History
mode.
Home
“Exposure Access keys”
pressed when History
preset mode is selected.
ATTENTION
Press UP/DOWN to
scroll, SET to accept or
MODE to cancel.
History/Mode Selection
Any one of the following
keys pressed.
1) SEL
2) EXPOSURE or
3) Exposure Access
ATTENTION
Press SET to accept
change or SEL to
discard.
Parameter Modification
Any one of the following
keys pressed.
1) Mode
2) EXPOSURE or
3) Exposure Access
ATTENTION
mA limited to 6.
Use mS < 210 for full
range. Press again to
continue.
Parameter Modification
- mA
UP / DOWN keys
pressed from 6mA / 4mA
respectively. 7mA-8mA
options not available
when mS is more than
200mS.
Содержание X-RAY
Страница 1: ...OPERATING MANUAL ...
Страница 3: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 37: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 45: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 57: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 59: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Страница 65: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...