When the selector knob is turned so the arrow is pointing from the left side to the right side, the set point of the right side regulator is increased 
to a pressure higher than the left side regulator.  Similarly, when the selector knob is turned so the arrow is pointing from the right side to the left 
side, the set point of the right manifold regulator is reduced to a pressure lower than the left side regulator. 

NOTE: Always move the selector 

knob gently but fully against the mechanical stop to prevent the off-line manifold from feeding unintentionally.

The regulator that is set to deliver the higher pressure will supply gas until its source is depleted, at which time the inlet pressure will decrease 
to just below the set point of the other regulator.  The other side will then be operating at a higher pressure, which will force the system to switch 
over and supply gas to the downstream process.

Model 920 Specifications

Maximum Rated Inlet Pressure –   

3000 psi

Outlet Pressure Ranges –   

10-100 psi

 

20-200 psi

Right Side Changeover Pressure 

280 psi for 200 psi model; 160 psi for 100 psi model

Left Side Changeover Pressure 

230 psi for 200 psi model; 140 psi for 100 psi model

Flow Coefficient -  

Cv = 0.03

Flow Capacity -  

200 scfh (air) @ 100 psi outlet

Design Leak Rate -  

2.0 x 10-6 cc/sec outboard He leak rate

Weight -  

 11 lbs.

Ambient Operating Temperature    

-40 F to 150 F

Inlet Connections -  

 Flexible Stainless Steel Corrugated Hose with CGA Check Valves

Outlet Port -  

 ¼” FNPT 

Materials

Body -  

 Brass or 316L SS

Bonnet -  

 Chrome Plated Brass Barstock 

Seat -  

 PTFE

Diaphragms -  

 316L SS

Gauges -  

 Brass or Stainless Steel

Filters -  

 10 Micron Plated Bronze or 316L SS

Содержание 920 Series

Страница 1: ...h guidelines stated by the National Fire Protection Association the Compressed Gas Association OSHA and all applicable local codes Switchover manifolds should not be placed in a location where the tem...

Страница 2: ...o not allow any oil grease or other potential contaminating lubricants to come in contact with any components These panels are factory assembled under rigorous high purity conditions without any use o...

Страница 3: ...below the set point of the other regulator The other side will then be operating at a higher pressure which will force the system to switch over and supply gas to the downstream process Model 920 Spec...

Страница 4: ...is properly cleaned to prevent the introduction of contaminants into the gas stream Harris Equipment Warranty Policy Harris warrants that manifold products sold in non corrosive gas service are free f...

Страница 5: ...ablecidas por la Asociaci n Nacional de Protecci n contra el Fuego la Compressed Gas Association OSHA y todos los c digos locales aplicables Los colectores de conmutaci n no deben colocarse en un luga...

Страница 6: ...xi n CGA de ajuste manual opcional No permita que ning n aceite grasa u otro lubricante contaminante entre en contacto con ning n componente Estos paneles se ensamblan en f brica en condiciones riguro...

Страница 7: ...ebajo del punto de ajuste del otro regulador El otro lado funcionar a una presi n m s alta lo que forzar al sistema a conmutar y suministrar gas al proceso posterior Especificaciones del modelo 920 Pr...

Страница 8: ...de gas Pol tica de garant a de equipos Harris Harris garantiza que los productos de colector vendidos para el servicio de gas no corrosivo est n libres de defectos de fabricaci n en materiales y mano...

Страница 9: ...a valve et le filetage du cylindre Consignes g n rales Les collecteurs commutateur doivent tre install s conform ment aux recommandations de la National Fire Protection Association la Compressed Gas A...

Страница 10: ...l ou la main en option Ne pas laisser d huile de graisse ou d autres lubrifiants potentiellement contaminants entrer en contact avec les composants Ces panneaux sont assembl s en usine dans des condit...

Страница 11: ...ndeur L autre c t fonctionnera alors une pression plus lev e ce qui forcera le syst me commuter et fournir du gaz au proc d en aval Sp cifications pour le mod le 920 Pression d admission nominale maxi...

Страница 12: ...la p n tration de contaminants dans le flux de gaz Politique de garantie de l quipement Harris Harris garantit que les collecteurs vendus pour le service au gaz non corrosif sont exempts de d fauts d...

Отзывы: