Hansen APP08 Скачать руководство пользователя страница 10

10

APP-001b
APR 2007

CONCENTRATING WATER mode.  This information is 
available for information only, and no action is required 

by the operator related to the counts.  The drain mode is 

always set to 0, and can not be changed.

To change the display units, move the cursor with the / \ 

and \ / keys to the units: # line and press enter.  Change 
the value with the / \ or \ / keys, and press enter again.  
Press esc to return to the menu.

The version menu item displays the software revisions 

for the purger’s display keypad and for the purger’s plc.

PLC MESSAGES

“Active Purge Point ##” where ## indicates the active 

purge point.

For 115VAC models, “Liquid Temp: ### ºF” will appear 
during the CONCENTRATING WATER process.

For 220VAC models, “Liquid Temp: ### ºC” will appear 
during the CONCENTRATING WATER process.

“READY  TO  DR AIN.  OPEN  WATER  PURGE  SHUT-OFF 

VALVE THEN PRESS “OK” BUTTON.” will appear when the 
process of CONCENTRATING WATER is complete and it 

is ready to be drained.

“DRAINING IS COMPLETE. CLOSE WATER PURGE SHUT-

OFF VALVE THEN PRESS “OK” BUTTON.” will appear when 

the APP is done draining and is ready to go back to its 
normal mode of COLLECTING AIR/ WATER.

“Releasing  Air”  will  appear  when  the  APP  is  releasing 

air.

“Concentrating Water” will appear during the process of 

heating the ammonia /water solution. 

CONTROL CONSOLE MENU HIERARCHY

The control console menu is as follows:

+air purge limit

 

60 minute limit: #

 

(normal=0, 60 min)

 (startup=1, 

nolimit)

+purge point enable

 01:# 

1=enable,0=skip

 02:# 

1=enable,0=skip

 03:# 

1=enable,0=skip

 04:# 

1=enable,0=skip

 05:# 

1=enable,0=skip

 06:# 

1=enable,0=skip

 07:# 

1=enable,0=skip

 08:# 

1=enable,0=skip

Model APP08 has purge points 01-08 as shown here, 
model APP16 has purge points 01-16, model APP24 has 
purge points 01-24.  Model APPC does not have this 
menu item.

+water collection

 counts= 

(600)

 

drain mode: 0

+display units

 units: 

#

 (1=US)

 (0=SI/metric)

+version

 

keypad FW= ####

 

plc FW= ####

To change the air purge limit, move the cursor with the / \

 

and \ / keys to the 60 minute limit: # line and press enter.  
Change the value with the / \ or \ / keys, and press enter 
again.  Press esc to return to the menu.

To enable or disable individual purge points, move the 

cursor with the / \ and \ / keys to the desired purge point 
and press enter.  Change the value with the / \ or \ / keys, 
and press enter again.  Press esc to return to the menu.  

At system start-up, when a purger is initially installed in 

a system that has fewer purge points than are available 
on the purger (e.g., an APP08 used on system that has 5 
purge points), the unused purge points must be set to 0.

The water collection parameters can not be changed and 

are for information only.  The counts indicate the time 

in seconds that the water concentration as measured 
by the water concentration sensor exceeds the 

concentration set point.  The concentration of water in 
the purger evaporator must exceed 25% for 10 minutes 
(600 seconds) continuously before the purge will enter 

Содержание APP08

Страница 1: ...RGER Plus is a totally automatic electronically controlled refrigerant noncondensible gas air and water purger for reducing the energy costs of operating an ammonia refrigeration system All models are...

Страница 2: ...circumstances only one point is effectively purged The best practice is to purge each condenser and receiver circuit separately It is extremely important to install purge points at locations sure to b...

Страница 3: ...some cases a small high pressure auxiliary receiver is located at the outlet of one or more condensers This receiver should have a purge point at the top Where a high pressure float regulator is used...

Страница 4: ...ected into the purger The low pressure pumped liquid line feeds makeup liquid as required during purging The line contains liquid ammonia that typically is also contaminated with water which is to be...

Страница 5: ...mps electrical supply on a 15 amp circuit breaker are available All models have a 13 mm knockout on the side of the control cabinet to access the power connection terminal strip Any unused knockout ho...

Страница 6: ...connecting shut off valves are open See Figure 4 START UP Make sure all piping electrical connections and settings are complete as described in this bulletin Open the foul gas low pressure pumped liq...

Страница 7: ...there are no releases to atmosphere 1 WATER BUBBLER WATER BUBBLER OVERFLOW TUBE FOUL GAS LINE LOW PRESSURE PUMPED LIQUID LINE SUCTION LINE FILL LINE D B C FLOW REGULATOR GAUGE VALVE GAUGE VALVE LIQUI...

Страница 8: ...system parameters The operator can switch between the two message types by pressing the esc key Control Console Specifications Display Type Character LCD 4 lines by 20 characters w LED backlight Encl...

Страница 9: ...and the temperature of the plate heaters see figure 4 Another press of this key displays the pressures in the low side and the high side of the purger A final press of this key returns the display to...

Страница 10: ...follows air purge limit 60 minute limit normal 0 60 min startup 1 nolimit purge point enable 01 1 enable 0 skip 02 1 enable 0 skip 03 1 enable 0 skip 04 1 enable 0 skip 05 1 enable 0 skip 06 1 enable...

Страница 11: ...normal 7777 loss of lp liq will appear when the liquid make up solenoid valve has been energized for 30 minutes continuously ERROR HEATER FAULT will be displayed if the heaters are on longer than 4 ho...

Страница 12: ...1 psid 07 bar check valve in the water purge line to prevent reverse flow of water into the purger from the customer supplied container The water bubbler fill line has a check valve OPERATION OF PURG...

Страница 13: ...URGE GAS GAUGE VALVE WIDE OPEN DURING OPERATION CLOSED FOR SERVICING PURGER PURGE GAS 1 32 0 8 MM DIA METERING ORIFICE 0 07 BAR PURGE GAS TO WATER BUBBLER OR TO CUSTOMER SUPPLIED WATER BOTTLE CHECK VA...

Страница 14: ...r fill line valve The purger will pump down in several hours To accelerate the process attach ammonia hoses to the oil drain valves and pump out into a suction line Close the low pressure liquid retur...

Страница 15: ...por Trap H5600R Relief Valve 300 psi ELECTRICAL 70 1088 Solenoid Coil DIN plug 115V 50 60 Hz 16 Watt 20 2428 Electric Heater 115V 50 60 Hz 750 Watt 20 2477 Electric Heater 230V 50 60 Hz 750 Watt 20 12...

Страница 16: ...ctional pushbutton several times until the pressures are displayed Verify that the high side pressure is greater than 80 psig 5 5 barg Action If the high side pressure displayed is less than 80 psig 5...

Страница 17: ...id line plugged strainers or the plastic shipping cap still in the liquid line flange Action Inspect piping for a closed shut off valve Open the shut off valves clean the strainer or remove the shippi...

Страница 18: ...18 APP 001b APR 2007 THIS PAGE INTENTIONALLY BLANK...

Страница 19: ...ce schedule and the Safety Precautions Sheet supplied with this product Escaping refrigerant can cause injury especially to the eyes and lungs WARRANTY All Hansen products except electronics are guara...

Страница 20: ...poration 6827 High Grove Boulevard Burr Ridge Illinois 60527 USA Tel 630 325 1565 Fax 630 325 1572 Toll 800 426 7368 Email info hantech com Web www hantech com USA Asia Europe India Latin America Midd...

Отзывы: