19. TROUBLESHOOTING GUIDE
SYMPTOMS
PROBLEM
SOLUTION
Display shows DO reading
blinking.
Reading out of range.
Recalibrate the meter;
Check the sample is within
measurable range.
Meter shuts off.
Dead batteries; Auto Power Off
feature is enabled: in this case,
meter shuts off after selected
period of non use.
Replace batteries; Press
ON/OFF
.
The meter does not start when
pressing
ON/OFF
.
Initialization error.
Press and hold down
ON/OFF
for
about 20 seconds or disconnect
and then connect the batteries.
Display shows “
No Cap
Detected
”.
Cap not on correctly.
Check Cap/Reseat Cap.
Display shows “
No Cap Info
Detected
” with Cap.
Cannot read Cap Info.
Check Cap and Retry. Replace Cap
if it continues.
Display shows “Cap Damaged”.
Sensor Damaged.
Try reseating cap. If this doesn’t
resolve issue, replace Cap.
Display shows “
Cap Expired
”.
Cap reached lifetime use
expiration.
Continue OR Replace Cap
(Continuing use may result in
incorrect measurements).
Display “
No Probe
”.
Probe not present/ Not connected
properly.
Connect/Disconnect and reconnect
probe OR Turn meter OFF then ON.
Meter shows “
Probe Err xx
”
message.
Probe Internal Error.
Disconnect and reconnect probe.
If problem persists replace the
probe.