19
SETUP
Highlight
OUR
configuration
and press
Select
.
Press / keys to modify the selected parameter’s value.
Press
Prev/Next
to select a different parameter.
Press
Save
to save the new OUR configuration.
Press
ESC
to leave without changing.
Parameters:
•
Min time
- the minimum time for the OUR test.
Range: 1 to 3600 seconds.
•
Max time
- the maximum time for the OUR test. The test will stop automatically when the
maximum time has elapsed.
Range: 1 to 3600 seconds.
•
Min start DO
- the minimum accepted DO value for starting the OUR test. If the DO reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
•
Min end DO
- the minimum accepted DO value at the end of the test. If the DO reading at the
end of the OUR test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
•
Total volume
- the volume of the diluted mixture.
Range: 0.1 to 300.0 mL
•
Sample volume
- the volume of sample in the diluted mixture.
Range: 0.1 to 300.0 mL.
SOUR Configuration
Specific Oxygen Uptake Rate (SOUR), also known as the oxygen consumption or respiration rate, is
defined as the milligram of oxygen consumed per gram of volatile suspended solids (VSS) per hour.
These parameters will be used to flag error limits and make dilution calculations for the determination.
Ignore this parameter if not making SOUR measurements.