45
OUR MEASUREMENT
16. OUR MEASUREMENT
The
OUR is used to determine the oxygen consumption or respiration rate in water. It is defined as the
mg/L of oxygen consumed per hour.
The following equation is used for
OUR determination:
where:
DO
START
=
Dissolved oxygen level at start of test
DO
END
=
Dissolved oxygen level at end of test
t
ELAPSED
=
Elapsed time of test in seconds
total volume/sample volume = Dilution factor of sample
Before starting an OUR test remember to set the OUR configuration from the
Setup
menu.
This a an example of possible setting. Follow your normal SOP.
Make sure the
DO
probe has been calibrated before using. Clean off probe between samples.
Press
MODE
to select the
OUR measurement mode. Place the calibrated probe into the airtight
container of sample. Stir the sample. Press
Start
to begin taking measurements.
During the analysis the meter will display the ongoing calculation.
OUR recall: